pipet4u pro manual pipettes
Fixed and variable volume, with a glove-friendly counter lock.
AHN pipet4u pro: single-, 8- or 12-channel air-displacement pipettes, from 0.1 µL to 10 mL depending on version.
Reverse pipetting, blow-out and universal tip compatibility suit routine laboratory liquid handling.
| Mechanism | Air displacement |
|---|---|
| Channels | 1, 8 or 12; no 16-channel version |
| Volume type | Fixed or variable |
| Overall range | Fixed 5 µL–10 mL; variable 0.1–2.5 µL through 2–10 mL |
| Single-channel fixed volume (volume; systematic; random error) | 5 µL — 1,30 % — 1,20 %; 10 µL — 0,80 % — 0,80 %; 20 µL — 0,60 % — 0,50 %; 25 and 50 µL — 0,50 % — 0,30 %; 100 µL — 0,50 % — 0,30 %; 200 and 250 µL — 0,40 % — 0,20 %; 500 and 1000 µL — 0,30 % — 0,20 %; 2, 2,5 and 5 mL — 0,30 % — 0,15 %; 10 mL — 0,60 % — 0,20 % |
| Single-channel variable volume (volume; increment; systematic and random error at three points) | 0,1–2,5 µL — 0,01 µL — 12,00/3,00/2,50 % — 6,00/3,00/1,60 %; 0,5–10 µL — 0,10 µL — 2,50/2,00/1,00 % — 1,50/1,00/0,80 %; 2–20 µL — 0,10 µL — 3,00/1,50/0,90 % — 2,00/1,00/0,40 %; 5–50 µL — 0,50 µL — 2,00/0,80/0,60 % — 2,00/0,40/0,30 %; 10–100 µL — 0,50 µL — 3,00/1,00/0,80 % — 1,50/0,50/0,15 %; 20–200 µL — 1,00 µL — 2,00/0,70/0,60 % — 0,80/0,30/0,15 %; 50–200 µL — 1,00 µL — 1,00/0,80/0,60 % — 0,40/0,20/0,15 %; 100–1000 µL — 5,00 µL — 2,00/1,00/0,60 % — 0,70/0,40/0,20 %; 200–1000 µL — 5,00 µL — 0,90/0,75/0,60 % — 0,30/0,25/0,20 %; 0,5–5 mL — 0,05 mL — 2,00/0,60/0,50 % — 0,60/0,30/0,15 %; 2–10 mL — 0,10 mL — 3,00/1,20/0,60 % — 0,60/0,30/0,20 % |
| 8/12-channel variable volume | 0,5–10 µL — 0,10 µL — 4,00/2,50/1,50 % — 4,00/2,50/1,50 %; 5–50 µL — 0,50 µL — 3,00/1,50/1,00 % — 2,00/1,00/0,70 %; 10–100 µL — 0,50 or 1,00 µL — 1,50/1,00/1,00 % — 0,75/0,50/0,25 %; 30–300 µL — 1,00 µL — 1,50/1,00/0,80 % — 0,75/0,50/0,25 % |
| Reverse pipetting | Supported |
| Blow-out | Provided for complete dispensing |
| Volume lock | Glove-friendly counter lock |
| Identification | Exchangeable colored markers with installation tool |
| Finger rest | Soft TPE |
| Tip compatibility | Universal, including common third-party brands |
| Autoclaving | Whole pipette at 121 °C for 15 min; calibration retained after repeated cycles |
| Standards | DIN EN ISO 8655-1, -2 and -6 |
| Calibration certificate | Included |
| Warranty | Two years |
| Supplied items | Pipette, calibration/disassembly tool, colored markers and tool, silicone grease, sample tips, O-rings, manual, warranty card and calibration report |