contrAA 800 (HR-CS AAS)
One continuous lamp for every element with access to any absorption line and its spectral neighborhood.
contrAA 800 is high-resolution continuum-source AAS with a xenon lamp, echelle optics and 200-pixel CCD over 185–900 nm for up to 72 elements.
contrAA 800 is high-resolution continuum-source AAS (HR-CS AAS). One xenon short-arc lamp replaces element-specific hollow-cathode lamps over 185–900 nm, opening alternate lines for up to 72 elements including nonmetals.
A fast sequential 380 mm echelle and quartz-prism/grating double monochromator feed a 200-pixel CCD. It measures the analytical line and its neighboring spectrum, allowing structured-background correction after acquisition. F offers flame/hydride, G graphite furnace/hydride, and D both with automatic atomizer switching. Solid AA and HydrEA are optional.
| Instrument type | High-resolution continuum-source atomic absorption spectrometer (HR-CS AAS) |
|---|---|
| Versions | 800 F flame/hydride; 800 G graphite furnace/hydride; 800 D flame/furnace/hydride |
| Source | Water-cooled xenon short-arc continuum lamp, replaceable without alignment |
| Spectral range | 185–900 nm |
| Monochromator | Fast sequential 380 mm echelle, single-beam; quartz-prism/rotating-grating double monochromator |
| Spectral resolution | 0.002 nm at 200 nm (As 193.7 nm line FWHM ≤3.8 pm) |
| Wavelength repeatability | 0.4 pm |
| Detector | 200-pixel CCD line array |
| Elements | Up to 72 including nonmetals |
| Concentration range | Fractions of µg/L to percent |
| Atomization | Flame with coded titanium burner; transversely heated graphite/STPF; hydride/cold vapor; optional Solid AA/HydrEA |
| Background correction | Iterative baseline (IBC), dynamic/static, structured-background reference and least-squares (LSBC) |
| Flame sensitivity | Cu 324 nm: 0.015 mg/L at 1% absorbance, 100 mm burner |
| Graphite sensitivity | Pb 283 nm: 0.66 µg/L at 1% absorbance, 20 µL peak area |
| Dimensions (W×D×H) | 780 × 775 × 625 mm |
| Weight | 140 kg (F); 170 kg (G/D) |
| Power | 230/110 V (F); 230 V (G/D), 50/60 Hz |
| Power consumption | F 460 VA base/650 VA max; G/D 2100 VA base/20,450 VA max |
| Gases | Acetylene 2.6 at 80–160 kPa; N₂O 2.5 at 400–600 kPa; oil-free air 400–600 kPa; Ar 4.8 at 600–700 kPa |
| Software | ASpect CS; optional 21 CFR Part 11 |