mercur DUO plus
Fluorescence for trace mercury and absorption for higher concentrations in one cold-vapor analyzer.
mercur DUO plus combines CV-AFS/CV-AAS at 253.7 nm with gold traps and six decades of dynamic range.
mercur DUO plus measures cold-vapor mercury by atomic fluorescence at 90° for trace levels and atomic absorption for higher levels, both at 253.7 nm. The combined six-decade range avoids changing instruments for different samples.
Gold amalgamation improves detection and separates Hg from other gases; dual traps meet EPA 1631 requirements. A 1 mL sample reacts in six seconds; a cycle without enrichment takes around 40 seconds. AS-F/AS-FD autosamplers offer up to 129 15 mL or 54 50 mL positions.
| Instrument type | Cold-vapor Hg analyzer with atomic fluorescence (CV-AFS) and absorption (CV-AAS) |
|---|---|
| Wavelength | 253.7 nm; single-beam DC method |
| Light source | UVU5 low-pressure Hg lamp with cup electrode |
| Detector | Nine-stage 1P28 photomultiplier |
| Fluorescence cell | Flow-through quartz, internal 10 × 10 × 32 mm |
| Dynamic range | Six orders of magnitude in ng/L and µg/L |
| Enrichment | Gold amalgamation; dual collectors meet EPA 1631 |
| Sample volume | 1 mL with 6 s reaction |
| Cycle time | About 40 s without enrichment |
| Separator | Low-dead-volume gas/liquid, reduced foaming |
| Autosamplers | AS-F/AS-FD: 129 × 15 mL, 10 × 50 mL or 54 × 50 mL |
| Dimensions | 600 × 350 × 490 mm |
| Weight | 37 kg |
| Gases | Hg-free Ar or He ≥99.999%, inlet ≤700 kPa, working 200 kPa, flow ≤60 L/h |
| Power | 230 V ±10%, 50/60 Hz; 90 VA average, 600 VA maximum |
| Operating conditions | +10 to +35 °C, ≤90% humidity at +30 °C |
| Software | WinAAS |
| Standard | EPA 1631 with dual gold collectors |
| Numerical detection limit | Not stated in available manufacturer documentation |