multi EA 5100
Carbon, nitrogen, sulfur and chlorine in liquids, solids, gases and LPG on one modular analyzer.
multi EA 5100 combines four-element combustion analysis, a furnace to 1100 °C and sulfur detection from 5 µg/L.
multi EA 5100 is a modular C/N/S/Cl microelemental analyzer across liquids, solids, gases and liquefied gases. Dedicated detection uses NDIR for carbon forms, chemiluminescence for nitrogen, UV fluorescence or coulometry for sulfur and coulometry for chlorine. Detection limits reach 5 µg/L S, 10 µg/L N, 50 µg/L Cl and 100 µg/L C.
Horizontal and vertical introduction supports 1–500 µL liquids, 0.1–110 mg solids, 1–100 mL gases and 1–50 µL LPG, with appropriate modules. Autosamplers, boat drive and gas/LPG modules are available. More than 100 prevalidated methods span ASTM, DIN EN, ISO, JIS, UOP and EPA.
| Instrument type | Modular combustion microelemental analyzer |
|---|---|
| Elements and parameters | C (TC, TOC, TIC, NPOC, EC, OC), N (TN), S (TS), Cl (TX, TOL, TCl) |
| Detectors | NDIR C; chemiluminescence N; UV fluorescence/coulometry S; coulometry Cl |
| Ranges | C to 100% or 10,000 mg/L in water (500 mg absolute); N to 10,000 mg/L (100 µg absolute); S to 10,000 mg/L UV or 40,000 mg/L coulometry; Cl to 100,000 mg/L (1 mg absolute) |
| Detection limits | S 5 µg/L (0.2 µg absolute); N 10 µg/L (0.4 ng); Cl 50 µg/L high-sensitivity (10 ng); C 100 µg/L (50 ng) |
| Furnace temperature | Up to 1100 °C |
| Sample introduction | Horizontal and vertical |
| Samples and volumes | Liquids 1–500 µL; solids 0.1–110 mg; gases 1–100 mL to 200 bar; LPG 1–50 µL to 35 bar |
| Analysis time | 3–5 min vertical; 3–8 min horizontal or Cl |
| Throughput | Up to 160 liquid or 90 solid samples/day |
| Autosamplers/modules | AI/AI-EA injectors, MMS 5100/MMS-T, ABD boat drive, GSS gas, LPG 2.0 and GSS/LPG |
| Dimensions | Base 513 × 464 × 551 mm vertical; 1073 × 464 × 551 mm with boat drive |
| Weight | About 26 kg |
| Gases | Ar 99.996% (4.6) and O₂ 99.995% (4.5), halogen-/hydrocarbon-free |
| Power | 100–240 V AC, 50/60 Hz, up to 16 A |
| Software | multiWin 5.X with over 100 validated ready methods |
| Standards | S: ASTM D5453/D6667/D7183/D7551, DIN EN 17178/15486, ISO 20846, JIS K 2541-06, ASTM D3120/D3246, DIN EN ISO 16591; N: ASTM D5762/D4629/D6069/D7184, DIN 51444, JIS K 2609, UOP 936/971/981; Cl: ASTM D5808/D4929/D7457, EPA 9076, DIN EN 14077, UOP 779/910; C: VDI 2465, ISO 8245, DIN EN 1484; halogens: EPA 9020B/1650B/450.1, DIN EN ISO 9562, EN 16166 |