novAA 800
Routine AAS with RFID-recognized eight-lamp turret and flame, graphite or combined configurations.
novAA 800 uses hollow-cathode lamps, a Czerny-Turner monochromator and deuterium background correction.
novAA 800 is classical atomic absorption spectrometry for routine industrial QC. Its eight-position standard hollow-cathode-lamp turret recognizes lamps by RFID. A sealed purged Czerny-Turner monochromator uses a rotating 1800-lines/mm holographic grating and five automatic slit widths. A low-noise SiOSens solid-state detector reads linearly across 0.1–100%, with deuterium background correction.
F supports flame/hydride, G transversely heated graphite furnace, and D combines flame and furnace with automatic switching. Source, optics and detector are common across the range.
| Instrument type | Lamp-source atomic absorption spectrometer |
|---|---|
| Versions | 800 F flame/hydride; 800 G graphite; 800 D flame/graphite with automatic switch |
| Source | Eight-position RFID-recognized standard hollow-cathode lamps |
| Spectral range | 185–900 nm |
| Wavelength repeatability | 0.005 nm |
| Monochromator | Sealed purged Czerny-Turner, 350/389 mm focal length |
| Grating | Rotating holographic 1800 lines/mm |
| Slits | Automatic 0.2, 0.3, 0.5, 0.8, 1.2 nm |
| Detector | Low-noise SiOSens solid-state, linear 0.1–100% readout |
| Atomization | Flame; transversely heated THGA graphite/STPF; hydride and cold vapor |
| Background correction | Deuterium hollow-cathode lamp |
| Flame sensitivity | Cu 324 nm: 0.020 mg/L at 1% absorbance, 100 mm burner |
| Graphite sensitivity | Pb 283 nm: 0.80 µg/L at 1% absorbance, 20 µL peak area |
| Dimensions (W×D×H) | 820 × 770 × 600 mm |
| Weight | 95 kg (F); 125 kg (G); 130 kg (D) |
| Power | 230/110 V (F); 230 V (G/D), 50/60 Hz |
| Fuse | 16 A (F); 35 A (G/D) |
| Gases | Acetylene 2.6 at 80–160 kPa; N₂O 2.5 at 400–600 kPa; air 400–600 kPa; Ar 4.8 at 600–700 kPa |