PlasmaQuant 9100
400 mm echelle optics over 160–900 nm provide more than 43,000 emission lines.
PlasmaQuant 9100 ICP-OES analyzes up to 75 elements with 40 MHz plasma and Elite resolution to 0.002 nm.
PlasmaQuant 9100 is ICP-OES for up to 75 elements in complex liquid matrices. Its 400 mm double echelle with quartz-prism premonochromator covers 160–900 nm and more than 43,000 lines to work around overlaps. Elite resolution is 0.002 nm at 200 nm versus 0.006 nm standard.
A 40 MHz, 700–1700 W free-running tube generator supplies plasma. Radial, axial and attenuated Dual View PLUS observation are available. Argon flow stops in standby. Applications include chemicals, oil/gas, mining and metals.
| Instrument type | Inductively coupled plasma optical emission spectrometer (ICP-OES) |
|---|---|
| Versions | PlasmaQuant 9100 and 9100 Elite |
| Elements | Up to 75 |
| Optics | Double echelle with quartz-prism premonochromator, 76° blaze grating, 400 mm focal length |
| Spectral range | 160–900 nm; more than 43,000 emission lines |
| Resolution | Standard 0.006 nm at 200 nm (FWHM ≤5.0 pm); Elite 0.002 nm (FWHM ≤3.5 pm) |
| Detector | CCD cooled to −10 °C; six-decade linear range; 1 ms–10 s integration |
| Plasma generator | 40 MHz free-running tube, 700–1700 W in 50 W steps, four-turn copper coil |
| Plasma view | Radial, axial and attenuated Dual View PLUS |
| Argon use | Plasma 10–20 L/min; auxiliary 0.2–2.0; nebulizer 0.1–1.5; O₂ option 0–0.05; total 13–21; 4–6 bar inlet; none at standby |
| Detection limits in 0.5% HNO₃ | P 177.436 <2.0; As 193.698 <2.0; Zn 213.856 <0.1; Pb 220.353 <1.0; Mn 257.610 <0.05; V 292.401 <0.1; Cu 324.754 <0.2 µg/L |
| Dimensions (W×H×D) | 990 × 940 × 855 mm |
| Weight | About 170 kg |
| Power | 230 V ±10%, 50/60 Hz, 4600 VA, 32 A fuses |
| Cooling water | 1.5–2.0 L/min at 17–24 °C; exhaust 3.5–5.5 m³/min |
| Sample introduction | V Shuttle Torch, 12-roller peristaltic pump, Standard/Salt/HF/Organic kits |
| Software | ASpect PQ with QC and 21 CFR Part 11 support |