PlasmaQuant MS
Entry model with a robust plasma for high-salt matrices, organic solvents and petrochemical samples.
PlasmaQuant MS is a quadrupole ICP-MS with ReflexION ion optics, iCRC cell, argon consumption below 9 L/min and a 3–260 u mass range.
PlasmaQuant MS is a quadrupole ICP-MS designed for sensitive analysis of difficult matrices, including high dissolved solids, organic solvents and petrochemical feedstocks.
Eco Plasma technology combines an efficient RF system with a nearly center-grounded plasma, allowing plasma gas consumption below 9 L/min. The manufacturer claims up to 50% lower plasma operating costs. No plasma shield is used, and secondary discharge is avoided.
ReflexION optics bend the ion beam by 90° and focus it with a parabolic electrostatic field. The integrated iCRC collision/reaction cell removes interferences with helium and hydrogen; gas modes switch within three seconds. A discrete-dynode electron multiplier provides 11 orders of dynamic range. Uranium-238 sensitivity for the base model is 300 million counts per ppm.
| Instrument type | Quadrupole inductively coupled plasma mass spectrometer (ICP-MS) |
|---|---|
| Mass range | 3–260 u with zero-blast protection |
| Resolution | Adjustable, 0.5–1.2 u |
| Quadrupole scan speed | 5,115 u/s; minimum integration time 50 µs |
| Mass calibration stability | 0.05 u per day |
| Sensitivity | Lithium-7: 65; beryllium-9: 20; cobalt-59: 400; indium-115: 500; lead-208: 225; thorium-232: 300; uranium-238: 300 million counts/ppm |
| Low-mass background (5 u) | Below 0.5 counts per second |
| Detection limits | Beryllium-9: 1.2; cobalt-59: 0.2; indium-115: 0.03; bismuth-209: 0.03; uranium-238: 0.02 ng/L |
| Plasma generator | Solid-state, 27 MHz, 300 V, 300–1600 W in 10 W steps, no plasma shield |
| Argon consumption | Plasma 7.5–10.5 L/min; auxiliary 0.85–1.85 L/min; nebulizer 1 L/min; purity at least 4.6 |
| Ion optics | ReflexION: 90° beam deflection, segmented ion mirror, three extraction lenses |
| Interference removal | Integrated iCRC collision/reaction cell with He and H2, BOOST technology, mode switching within 3 s |
| Detector | Discrete-dynode electron multiplier, 11 orders of dynamic range (0.1–10¹⁰ counts per second) |
| Cones | High-efficiency nickel; optional platinum tip |
| Backing pump | Pfeiffer SmartVane55 |
| CeO/Ce oxide ions | Below 2%; doubly charged Ba2+/Ba+: below 3% |
| Dimensions (W×D×H) | 660 × 589 × 1131 mm; weight 186 kg |
| Power | 200–240 V, 50/60 Hz, 25 A fuse, typical consumption 2,700 W |
| Software | ASpect MS; optional 21 CFR Part 11 and ASpect NP for nanoparticle analysis |