PlasmaQuant MS Elite
Top-of-range ICP-MS with maximum sensitivity and full control of ion focusing for research.
PlasmaQuant MS Elite reaches 1,000 million uranium counts per ppm, includes platinum Elite Cones and detects uranium down to 0.005 ng/L.
PlasmaQuant MS Elite is the top research model, distinguished by maximum sensitivity, two plasma sample-introduction ports and full user control over ion focusing.
Uranium-238 sensitivity reaches 1,000 million counts per ppm, twice the Q and more than three times the base model; its detection limit is 0.005 ng/L. Platinum Elite Cones are included.
Two plasma introduction ports support specialized integrations such as laser ablation, nanoparticle detection and speciation analysis when an additional stream is needed. The shared platform retains a 27 MHz solid-state generator using less than 9 L/min of argon, ReflexION optics, an iCRC cell and an 11-order dynamic-range detector.
| Instrument type | Research-grade quadrupole inductively coupled plasma mass spectrometer (ICP-MS) |
|---|---|
| Mass range | 3–260 u with zero-blast protection |
| Resolution | Adjustable, 0.5–1.2 u |
| Sensitivity | Lithium-7: 250; beryllium-9: 50; cobalt-59: 1,100; indium-115: 1,500; lead-208: 580; thorium-232: 1,000; uranium-238: 1,000 million counts/ppm |
| Low-mass background (5 u) | Below 1 count per second |
| Detection limits | Beryllium-9: 0.4; cobalt-59: 0.07; indium-115: 0.01; bismuth-209: 0.01; uranium-238: 0.005 ng/L |
| Configuration features | Two plasma introduction ports; full control of ion focusing |
| Plasma generator | Solid-state, 27 MHz, 300–1600 W in 10 W steps |
| Argon consumption | Plasma 7.5–10.5 L/min; auxiliary 0.85–1.85 L/min; nebulizer 1 L/min |
| Ion optics | ReflexION with 90° beam deflection |
| Interference removal | iCRC collision/reaction cell with He and H2; mode switching within 3 s |
| Detector | Discrete-dynode electron multiplier; 11 orders of dynamic range |
| Cones | Platinum Elite Cones; nickel available |
| Backing pump | Oil-free Edwards XDS 46 or Kashiyama NeoDry36E |
| Isotope-ratio precision | Boron-10/11 below 0.06% RSD; strontium-87/86 below 0.01%; lead-206/207 below 0.02%; uranium-235/238 below 0.07% |
| Dimensions (W×D×H) | 660 × 589 × 1131 mm; weight 186 kg |
| Software | ASpect MS; optional 21 CFR Part 11 and ASpect NP for nanoparticle analysis |