PlasmaQuant MS Elite S
Optimized signal-to-noise ratio for ultratrace analysis and purity control, including semiconductors.
PlasmaQuant MS Elite S combines 700 million uranium counts per ppm with low background for ultratrace and purity testing.
PlasmaQuant MS Elite S is a specialized model combining high sensitivity with the range's best signal-to-noise ratio. Its target applications are ultratrace analysis and material purity control, including semiconductor manufacturing.
Uranium-238 sensitivity of 700 million counts per ppm falls between the Q and Elite. Low-mass background is below 0.7 counts per second, as on the Q and below the Elite; the uranium-238 detection limit is 0.006 ng/L.
It uses the shared 27 MHz solid-state Eco Plasma platform with less than 9 L/min argon consumption, ReflexION ion optics, an iCRC helium/hydrogen collision/reaction cell and an 11-order dynamic-range detector. Pro Cones and an oil-free backing pump are included.
| Instrument type | Quadrupole ICP-MS for ultratrace analysis |
|---|---|
| Mass range | 3–260 u with zero-blast protection |
| Resolution | Adjustable, 0.5–1.2 u |
| Sensitivity | Lithium-7: 150; beryllium-9: 35; cobalt-59: 800; indium-115: 1,100; lead-208: 450; thorium-232: 700; uranium-238: 700 million counts/ppm |
| Low-mass background (5 u) | Below 0.7 counts per second |
| Detection limits | Beryllium-9: 0.5; cobalt-59: 0.09; indium-115: 0.01; bismuth-209: 0.01; uranium-238: 0.006 ng/L |
| Plasma generator | Solid-state, 27 MHz, 300–1600 W in 10 W steps |
| Argon consumption | Plasma 7.5–10.5 L/min; auxiliary 0.85–1.85 L/min; nebulizer 1 L/min |
| Ion optics | ReflexION with 90° beam deflection |
| Interference removal | iCRC collision/reaction cell with He and H2; mode switching within 3 s |
| Detector | Discrete-dynode electron multiplier; 11 orders of dynamic range |
| Cones | Nickel Pro Cones; optional platinum |
| Backing pump | Oil-free Edwards XDS 46 or Kashiyama NeoDry36E |
| Dimensions (W×D×H) | 660 × 589 × 1131 mm; weight 186 kg |
| Power | 200–240 V, 50/60 Hz, 25 A fuse |
| Software | ASpect MS; optional 21 CFR Part 11 and ASpect NP |