PlasmaQuant MS Q
Routine high-throughput ICP-MS: more than 80 water samples per hour at about 2% precision.
PlasmaQuant MS Q processes over 80 water samples per hour and detects uranium down to 0.009 ng/L.
PlasmaQuant MS Q is the range's versatile high-throughput model for laboratories processing many similar samples, including environmental monitoring, drinking and wastewater testing, and routine production control.
It processes more than 80 water samples per hour at approximately 2% relative standard deviation. Uranium-238 sensitivity is 500 million counts per ppm versus 300 for the base model, and its detection limit is 0.009 ng/L.
The shared platform includes a 27 MHz solid-state Eco Plasma generator using less than 9 L/min of argon, ReflexION ion optics, an iCRC helium/hydrogen collision/reaction cell and an 11-order dynamic-range detector. Compared with the base model, the Q uses Pro Cones and a dry backing pump.
| Instrument type | Quadrupole inductively coupled plasma mass spectrometer (ICP-MS) |
|---|---|
| Mass range | 3–260 u with zero-blast protection |
| Resolution | Adjustable, 0.5–1.2 u |
| Quadrupole scan speed | 5,115 u/s; minimum integration time 50 µs |
| Sensitivity | Lithium-7: 100; beryllium-9: 25; cobalt-59: 600; indium-115: 800; lead-208: 315; thorium-232: 500; uranium-238: 500 million counts/ppm |
| Low-mass background (5 u) | Below 0.7 counts per second |
| Detection limits | Beryllium-9: 0.8; cobalt-59: 0.1; indium-115: 0.02; bismuth-209: 0.02; uranium-238: 0.009 ng/L |
| Throughput | More than 80 water samples per hour at about 2% RSD |
| Plasma generator | Solid-state, 27 MHz, 300–1600 W in 10 W steps, no plasma shield |
| Argon consumption | Plasma 7.5–10.5 L/min; auxiliary 0.85–1.85 L/min; nebulizer 1 L/min |
| Ion optics | ReflexION with 90° beam deflection |
| Interference removal | Integrated iCRC collision/reaction cell with He and H2; mode switching within 3 s |
| Detector | Discrete-dynode electron multiplier; 11 orders of dynamic range |
| Cones | Nickel Pro Cones; optional platinum |
| Backing pump | Edwards XDS 46 or Kashiyama NeoDry36E |
| Dimensions (W×D×H) | 660 × 589 × 1131 mm; weight 186 kg |
| Power | 200–240 V, 50/60 Hz, 25 A fuse |
| Software | ASpect MS; optional 21 CFR Part 11 and ASpect NP |