SPECORD 200 PLUS
Classic full double-beam UV-Vis for daily work with a fixed 1.4 nm slit.
SPECORD 200 PLUS spans 190–1100 nm, scans up to 12,000 nm/min and reaches ±0.003 A visible accuracy.
SPECORD 200 PLUS is a classic Analytik Jena double-beam UV-Vis with a single monochromator and fixed 1.4 nm slit. Unlike Split Beam 50 PLUS, fully separate sample and reference paths compensate long-term drift, useful for kinetics and dissolution testing.
It shares 190–1100 nm range, ±0.1 nm wavelength accuracy, up to 12,000 nm/min scanning, ±0.003 A visible photometric accuracy and at most 0.0001 A baseline noise with the 50 PLUS. Two uncooled photodiodes detect light. Choose 210 PLUS for adjustable slits, 1200 nm reach and cooled detectors, or 250 PLUS for the lowest stray light from a double monochromator.
USP/Ph. Eur. validation and a 21 CFR Part 11 module are optional. It accepts the series' cell-compartment accessories.
| Instrument type | Full double-beam UV-Vis with fixed slit and single monochromator |
|---|---|
| Wavelength range | 190–1100 nm |
| Spectral slit width | Fixed 1.4 nm |
| Wavelength accuracy | ±0.1 nm at 656.1 nm; ±0.5 nm at 360.9 nm using holmium oxide |
| Wavelength repeatability | At most 0.02 nm RMS at 360.9 nm |
| Scan speed | Up to 12,000 nm/min |
| Minimum integration time | 0.001 s; minimum data step 0.02 nm |
| Photometric range | −3 to 3 A; software display −9 to 9 A |
| Measurement modes | Energy, absorbance, transmittance and reflectance |
| Photometric accuracy | ±0.003 A visible; ±0.010 A UV |
| Photometric repeatability | At most 0.0005 |
| Baseline noise at 500 nm | At most 0.0001 A RMS |
| Baseline deviation | ±0.0005 A; long-term stability at 500 nm ±0.0005 A/h |
| Zero transmittance | ±0.05% T |
| Stray light | At most 0.3% T at 198 nm (KCl); 0.03% T at 220/240 nm (NaI); 0.02% T at 340 nm (NaNO₂) |
| Light sources | Halogen and deuterium lamps; adjustable crossover 300–450 nm |
| Detectors | Two uncooled photodiodes |
| Optics | Single monochromator with imaging grating and quartz-coated aspheric optics |
| Internal calibration | Holmium oxide filter |
| Cuvette compartment (W×D×H) | 364 × 185 × 200 mm |
| Software | ASpect UV |
| Pharmacopoeia support | Optional validation modules for USP, Ph. Eur. and Analytik Jena specifications |
| 21 CFR Part 11 | Optional FDA 21 CFR Part 11 software module |
| Certification | ISO 9001, CE marking |
| Dimensions (W×D×H) | 590 × 690 × 290 mm |
| Weight | 22 kg |
| Operating conditions | 15–35 °C; relative humidity up to 90% at 30 °C |
| Power | 85–264 V AC, 50–60 Hz |