SPECORD 210 PLUS
The series’ broadest 185–1200 nm range with 0.2–4 nm variable slit and Peltier-cooled detectors.
SPECORD 210 PLUS combines 185–1200 nm reach, 0.2/0.5/1/2/4 nm slits and automatic wavelength calibration.
SPECORD 210 PLUS is a double-beam UV-Vis with variable slit and the broadest wavelength range in the series. Its 0.2, 0.5, 1, 2 and 4 nm slits allow narrow spectral lines, rare-earth solutions, solids and resolution-controlled pharmacopoeial methods while wider settings improve weak-signal throughput.
Only this model spans both deep UV to 185 nm and near-IR to 1200 nm. Peltier-cooled photodiodes reduce dark-current drift at range limits, and wavelength calibration is automatic. Stray light at 340 nm is at most 0.01% T, half that of 50/200 PLUS.
Like the range, it scans to 12,000 nm/min with ±0.003 A visible accuracy and at most 0.0001 A baseline noise. USP/Ph. Eur. validation and 21 CFR Part 11 are optional.
| Instrument type | Double-beam UV-Vis with variable slit and single monochromator |
|---|---|
| Wavelength range | 185–1200 nm, broadest in series |
| Spectral slit width | Variable 0.2, 0.5, 1, 2 or 4 nm |
| Wavelength accuracy | ±0.1 nm at 656.1 nm; ±0.5 nm at 360.9 nm using holmium oxide |
| Wavelength repeatability | At most 0.02 nm RMS at 360.9 nm |
| Scan speed | Up to 12,000 nm/min |
| Minimum integration time | 0.001 s; minimum data step 0.02 nm |
| Photometric range | −3 to 3 A; software display −9 to 9 A |
| Measurement modes | Energy, absorbance, transmittance and reflectance |
| Photometric accuracy | ±0.003 A visible; ±0.010 A UV |
| Photometric repeatability | At most 0.0005 |
| Baseline noise at 500 nm | At most 0.0001 A RMS |
| Baseline deviation | ±0.0005 A; long-term stability at 500 nm ±0.0005 A/h |
| Zero transmittance | ±0.05% T |
| Stray light | At most 0.3% T at 198 nm (KCl); 0.03% T at 220/240 nm (NaI); 0.01% T at 340 nm (NaNO₂) |
| Light sources | Halogen and deuterium lamps; adjustable crossover 300–450 nm |
| Detectors | Two Peltier-cooled photodiodes |
| Optics | Single monochromator with imaging grating and quartz-coated aspheric optics |
| Calibration | Holmium oxide filter; automatic wavelength calibration |
| Cuvette compartment (W×D×H) | 364 × 185 × 200 mm |
| Software | ASpect UV |
| Pharmacopoeia support | Optional validation modules for USP, Ph. Eur. and Analytik Jena specifications |
| 21 CFR Part 11 | Optional FDA 21 CFR Part 11 software module |
| Certification | ISO 9001, CE marking |
| Dimensions (W×D×H) | 590 × 690 × 290 mm |
| Weight | 22 kg |
| Operating conditions | 15–35 °C; relative humidity up to 90% at 30 °C |
| Power | 85–264 V AC, 50–60 Hz |