SPECORD 250 PLUS
Double monochromator cuts stray light to 0.005% T for turbid and strongly absorbing samples.
SPECORD 250 PLUS has the series’ double monochromator, 0.2–4 nm variable slit and Peltier-cooled detectors.
SPECORD 250 PLUS is the top Analytik Jena model and the only one in the range with a double monochromator. A second dispersion stage rejects light outside the selected wavelength, reducing stray light to 0.005% T at 220, 240 and 340 nm versus 0.02–0.03% T on single-monochromator models. This improves measurements above 2–3 A and on turbid samples.
Its variable slit covers 0.2–4 nm and photodiodes are Peltier cooled. The 190–1100 nm range is narrower than the 210 PLUS (185–1200 nm); select by required wavelength coverage and stray-light level. Other series metrology applies. Optional USP/Ph. Eur. validation, a USP <857> technical note and 21 CFR Part 11 module are available.
| Instrument type | Double-beam UV-Vis with variable slit and double monochromator |
|---|---|
| Wavelength range | 190–1100 nm, narrower than 210 PLUS |
| Spectral slit width | Variable 0.2, 0.5, 1, 2 or 4 nm |
| Wavelength accuracy | ±0.1 nm at 656.1 nm; ±0.5 nm at 360.9 nm using holmium oxide |
| Wavelength repeatability | At most 0.02 nm RMS at 360.9 nm |
| Scan speed | Up to 12,000 nm/min |
| Minimum integration time | 0.001 s; minimum data step 0.02 nm |
| Photometric range | −3 to 3 A; software display −9 to 9 A |
| Measurement modes | Energy, absorbance, transmittance and reflectance |
| Photometric accuracy | ±0.003 A visible; ±0.010 A UV |
| Photometric repeatability | At most 0.0005 |
| Baseline noise at 500 nm | At most 0.0001 A RMS |
| Baseline deviation | ±0.0005 A; long-term stability at 500 nm ±0.0005 A/h |
| Zero transmittance | ±0.05% T |
| Stray light | At most 0.03% T at 198 nm (KCl); 0.005% T at 220, 240 and 340 nm |
| Light sources | Halogen and deuterium lamps; adjustable crossover 300–450 nm |
| Detectors | Two Peltier-cooled photodiodes |
| Optics | Double monochromator with imaging grating and quartz-coated aspheric optics |
| Internal calibration | Holmium oxide filter |
| Cuvette compartment (W×D×H) | 364 × 185 × 200 mm |
| Software | ASpect UV |
| Pharmacopoeia support | Optional USP, Ph. Eur. and Analytik Jena validation modules; separate USP <857> technical note |
| 21 CFR Part 11 | Optional FDA 21 CFR Part 11 software module |
| Certification | ISO 9001, CE marking |
| Dimensions (W×D×H) | 590 × 690 × 290 mm |
| Weight | 23 kg |
| Operating conditions | 15–35 °C; relative humidity up to 90% at 30 °C |
| Power | 85–264 V AC, 50–60 Hz |