CONFOCHECK
Protein secondary structure in aqueous solution in about 30 seconds using a 7 µm AquaSpec flow cell.
Bruker CONFOCHECK measures aqueous proteins with below 0.1 mg/mL detection and 0–95 °C BioATR II temperature control.
CONFOCHECK is a dedicated Bruker FT-IR system for proteins in aqueous solutions. Water absorbs across the amide region used to determine secondary structure; the approximately 7 µm optical path of the AquaSpec flow cell allows the amide bands to be measured through water.
The system measures protein concentration, α-helices and β-sheets, conformational dynamics, thermal unfolding, aggregation, precipitation and crystallization. It also analyzes dissolved buffers, excipients and detergents.
BioATR II uses a silicon micro-ATR crystal with 0–95 °C temperature control, allowing continuous spectra during heating to locate denaturation. Each measurement takes about 30 seconds, with protein-concentration detection below 0.1 mg/mL. Stated service lives are ten years for laser and interferometer and five for the mid-IR source.
| Instrument type | Application-specific FT-IR system for aqueous proteins |
|---|---|
| Measurement time | About 30 s per sample |
| AquaSpec transmission cell | Flow cell, optimum optical path about 7 µm; for soluble and membrane proteins |
| Protein-concentration detection limit | Below 0.1 mg/mL |
| BioATR II accessory | Silicon-crystal micro-ATR, 0–95 °C control for thermal unfolding |
| Measured properties | Concentration; α-helix/β-sheet secondary structure; conformation and dynamics; unfolding, aggregation, precipitation and crystallization; aqueous buffers, excipients and detergents |
| Stated service life | Laser 10 years; interferometer 10 years; mid-IR source 5 years |