D8 DISCOVER
Versatile XRD for powders, thin films and epitaxial structures, with motorized beam geometry and samples to 50 kg.
D8 DISCOVER combines TRIO optics, TWIST-TUBE, TXS and IµS sources, EIGER2 R and LYNXEYE XE-T detectors.
D8 DISCOVER is a multipurpose diffractometer for powders, polycrystals, thin films, epitaxial structures and nanomaterials. Motorized TRIO optics switch between Bragg–Brentano, high-intensity parallel and high-resolution parallel beams without realignment. The optional ATLAS goniometer achieves ±0.007° 2θ accuracy over a range to 160°. A universal stage takes samples up to 50 kg and 300 mm diameter. Source and detector options support phase, texture, stress, reflectometry, grazing-incidence and scattering methods.
| Manufacturer | Bruker Corporation |
|---|---|
| Versions | D8 DISCOVER; DISCOVER Plus; DISCOVER HE (high energy, 2026) |
| Goniometer | Two-circle, independent stepper motors, optical encoders, maintenance-free drive |
| Optional ATLAS goniometer | Vertical, non-collinear arm; ±0.007° 2θ accuracy across range (NIST SRM 1976); 2θ to 160°; 0.001° minimum step |
| SOURCES | |
| TWIST-TUBE | Cr, Cu, Mo, Ag anodes; up to 3 kW; switchable 0.4 × 16 mm² point/line focus |
| Turbo X-ray Source (TXS/TXS-HE) | 0.3 × 3 mm² line focus; 6 kW/mm² focal brightness; Cu, Co, Cr, Mo anodes; 3.2 kW Cr, 5.4 kW Cu/Mo, 2.8 kW Co; up to 50 kV |
| IµS microfocus | Up to 50 W, single-phase; beam to 180 × 180 µm²; flux to 8 × 10⁸ counts/s at mirror exit; divergence to 0.5 mrad; MONTEL/MONTEL Plus optics |
| OPTICS | Motorized TRIO switches among three configurations: Bragg–Brentano; high-intensity Kα1,2 parallel beam; high-resolution Kα1 parallel beam. Ge(220)/Ge(004), two-/four-reflection Bartels monochromators; alignment-free SNAP.LOCK; PATHFINDER Plus |
| DETECTORS | EIGER2 R 250K/500K: up to 500,000 pixels, 75 × 75 mm², hybrid photon counting, 0D/1D/2D; LYNXEYE XE-T: <380 eV at 8 keV, Cr/Co/Cu/Mo/Ag anodes, no Kβ filters; manufacturer claims up to 450× conventional speed |
| STAGES | UMC: up to 50 kg and Ø300 mm, x/y ±150 mm, Z to 50 mm, unlimited φ, ψ to 55°; eucentric Eulerian cradle: ±40 mm, ψ −11° to 98°, sample to 1 kg |
| Nonambient chambers | 12–2500 K; 10⁻⁴ mbar–100 bar; 5–95% relative humidity |
| Methods | Qualitative/quantitative phase analysis; structure determination/refinement; microstrain/crystallite size; XRR; GID; in-plane/high-resolution diffraction; GISAXS; residual stress; texture/pole figures; microdiffraction; WAXS; PDF; SAXS |
| Software | DIFFRAC.SUITE: COMMANDER, WIZARD, EVA, LEPTOS, TEXTURE, TOPAS, SAXS |
| Dimensions, base version | 2020 × 1680 × 1290 mm (H × W × D) |
| Weight without optional electronics | 945 kg |
| Power supply | Single-phase 208–240 V or three-phase 120/230/240 V, 47–63 Hz |
| Generator power | Sealed tube 3 kW; IµS 50 W; TXS/TXS-HE 6 kW |
| Water cooling | External for sealed tube; none for IµS; external for Plus with TXS-HE |
| EU directives | 2014/30/EU EMC; 2014/35/EU electrical safety; 2006/42/EC machinery |