D8 QUEST
Everyday single-crystal chemical crystallography with entirely air-cooled operation.
D8 QUEST uses FIXED-CHI or KAPPA goniometry, microfocus sources and a PHOTON IV detector up to 140 cm².
D8 QUEST is a single-crystal diffractometer for routine chemical crystallography. Entirely air-cooled operation removes the need for an external water loop. FIXED-CHI and KAPPA four-axis goniometers provide a sphere of confusion below 7 µm. Sealed-tube and IµS microfocus sources use Cu, Mo or Ag anodes; PHOTON IV collects the diffraction data. Cryogenic attachments reach 28 K. Dual-source setups are directed to D8 VENTURE.
| Manufacturer | Bruker Corporation |
|---|---|
| Type | Single-crystal XRD for chemical crystallography |
| Goniometer | Four-axis FIXED-CHI or KAPPA; sphere of confusion <7 µm |
| Cooling | Entirely air cooled; no external water circuit |
| Sources | Sealed tube; IµS 3.0 microfocus; IµS DIAMOND II hybrid metal/diamond microfocus; Cu, Mo, Ag anodes |
| Two simultaneous sources | Unsupported; manufacturer directs dual-wavelength configurations to D8 VENTURE |
| Detector | PHOTON IV active area to 140 cm²; rare-earth scintillator with monolithic CMOS sensor; adjustable 2θ range and crystal-to-detector distance |
| Temperature attachments | Cryostream 800; Cryostream 1000/Plus; N-HELIX to 28 K |
| HTD high-temperature attachment to 1250 K | Manufacturer has not confirmed for QUEST; listed only for VENTURE |
| Software | APEX for chemical crystallography; PROTEUM |