minispec mq series (7.5 / 20 / 60 MHz)
Benchtop TD-NMR for production QC: fat, moisture, solid fat and fuel hydrogen by ISO, AOCS, IUPAC and ASTM methods.
Bruker minispec mq analyzers span 7.5, 20 and 60 MHz, use cryogen-free magnets and support sample temperatures from −100 to +200 °C.
Time-domain NMR measures free-induction or echo decay rather than a high-resolution spectrum. Signal behavior reveals the proton fraction in liquids and solids, enabling fast, nondestructive, solvent-free measurements of fat, moisture, solid content and hydrogen.
The minispec mq range includes mq7.5 for large samples, mq10 for 40 mm tubes, mq20 at 20 MHz with 18 mm tubes, and mq60 at 60 MHz. The platform spans 2–65 MHz and offers 0.17, 0.47 and 1.41 T fields. Lower frequency handles larger conductive or dense samples; higher frequency better separates components.
The permanent magnet needs no cryogens. Platform temperature control covers −100 to +200 °C, with narrower accessory ranges such as +5 to +50 °C for mq60 and −5 to +65 °C for mq20. Electronics include a 300 W transmitter, digital quadrature receiver, 20 ns pulse-programmer resolution and 12-bit, 32,000-point digitizer.
Applications include fat and moisture, solid fat content, fuel hydrogen, four-component polymer decay analysis and textile spin finish. Methods include AOCS Cd 16b-93, ISO 8292, IUPAC 2.150 and ASTM D7171/D3701/D4808. minispec Plus supports 21 CFR Part 11. The 75 × 65 cm platform weighs about 120 kg depending on configuration and operates on 100–240 V AC in rooms at 18–28 °C and 20–80% noncondensing humidity.
| Manufacturer | Bruker Corporation |
|---|---|
| Type | Time-domain NMR analyzers (TD-NMR) |
| Models | mq7.5 (7.5 MHz, large samples); mq10 (40 mm tubes); mq20 (20 MHz, 18 mm tubes); mq60 (60 MHz) |
| Overall platform frequency range | 2–65 MHz |
| Available field strengths | 0.17 T; 0.47 T; 1.41 T |
| Magnet | Cryogen-free permanent magnet; no liquid helium or nitrogen |
| Temperature control | Platform −100 to +200 °C; mq60 accessory +5 to +50 °C, mq20 accessory −5 to +65 °C |
| Transmitter | 300 W maximum with linear attenuation |
| Receiver | Digital phase-sensitive quadrature detection |
| Pulse programmer resolution | 20 ns |
| Digitizer | 12-bit, 32,000-point acquisition |
| Typical applications | Fat and moisture; solid fat content in fats and oils; hydrogen in fuels; four-component polymer decay analysis; textile spin finish |
| Standard methods (solid fat content) | AOCS Cd 16b-93; ISO 8292; IUPAC 2.150 |
| Standard methods (fuel hydrogen) | ASTM D7171; ASTM D3701; ASTM D4808 |
| Software | minispec Plus, 21 CFR Part 11 compatible |
| Platform dimensions | 75 × 65 cm (W×D) |
| Weight | About 120 kg depending on configuration |
| Power supply | 100–240 V AC, 50/60 Hz |
| Operating conditions | Room 18–28 °C; 20–80% noncondensing humidity |