Q4 TASMAN Series 2
Ten base metal matrices and three wavelength configurations tailored to the laboratory.
Q4 TASMAN Series 2 is a benchtop spark OES with ten matrices, 130/170/200 nm versions and pneumatic sample clamp.
Q4 TASMAN Series 2 is a benchtop spark spectrometer for broad metallurgical analysis. The 130 version spans 130–620 nm and measures vacuum-UV nitrogen and oxygen, including trace oxygen in pure copper. The 170 and 200 versions start at their named wavelengths and do not measure nitrogen.
MultiVision uses parallel Paschen-Runge optical paths: argon-purged 130–200 nm vacuum UV and unpurged 190–620 nm UV-visible. Thermally stabilized multichip detectors, a SmartSpark generator to 1000 Hz and a pneumatic stand for samples to 120 mm complete the platform.
| Instrument type | Benchtop spark OES |
|---|---|
| Optics | MultiVision dual parallel Paschen-Runge optical paths |
| Spectral range | Argon-purged vacuum UV 130–200 nm; unpurged UV-visible 190–620 nm |
| Versions | Q4 TASMAN 130 (130–620 nm), 170 (170–620 nm), 200 (200–620 nm) |
| Detectors | Thermally stabilized multichip systems |
| Base matrices | Ten: Fe, Al, Cu, Ni, Co, Ti, Pb, Sn, Zn, Mg |
| Nitrogen analysis | Only 130 version; best N/O performance including trace O in pure Cu |
| Excitation | Digital SmartSpark to 1000 Hz |
| Stand | Pneumatic clamp, samples to 120 mm high; coaxial argon flow |
| Dimensions (W×D×H) | 54 × 80 × 60 cm; up to 90 cm high with optional thermal module |
| Weight | About 75 kg |
| Power | 100–240 V ±10%, 50–60 Hz (manufacturer page says 100–204 V, likely typo); 16 A fuse at 240 V or 25 A at 100 V; 600 W measuring, 50 W standby |
| Software | ELEMENTAL.SUITE |