RAMANdrive
Raman and 325 nm photoluminescence for complete 300 mm semiconductor wafers and inspection coordinates.
RAMANdrive accepts 300 mm wafers on a 300 × 300 mm stage, acquires 400 spectra per pulse and supports dark-field and AreaFlash.
RAMANdrive Wafer Analysis System is Nanophoton/Bruker's semiconductor-focused Raman microscope. Its 300 × 300 × 35 mm stage holds whole 300 mm wafers with vacuum channels and locating pins. Defect coordinates from inspection equipment can be imported so the system automatically moves to each target.
A 325 nm UV laser supports photoluminescence mapping of defects, impurities, gallium distribution in GaN and InGaN quantum dots. A dark-field objective, unique in the range, detects particles below 100 nm. Line illumination records 400 spectra per pulse, while AreaFlash expands the beam for rapid large-area surveys. Confocal depth profiling maps stress in SiC/GaN and SiC polytypes.
Spectral resolution is better than 1.2 cm⁻¹, peak-position accuracy 0.1 cm⁻¹, and range starts at 100 cm⁻¹ (optional 50). Spatial resolution is 350 nm X, 500 nm Y and 1 µm Z with a 100×/0.90 NA objective. A 550 mm spectrometer, up to three gratings, cooled 1340 × 400 CCD and lasers from 325 to 785 nm support different tasks. The system weighs 390 kg including the vibration-isolation table.
| Manufacturer | Bruker Corporation (Nanophoton, a Bruker Company) |
|---|---|
| Full name | RAMANdrive Wafer Analysis System |
| Purpose | Semiconductor wafers up to 300 mm diameter |
| Stage | Motorized XYZ, 300 × 300 × 35 mm; vacuum channels and locating pins |
| Integration | Imports target coordinates from wafer-inspection equipment and moves automatically |
| Illumination | Line and point; 400 spectra per laser pulse |
| AreaFlash | Expanded laser spot for rapid single-image large-area overview |
| Photoluminescence | 325 nm UV laser for defects, impurities, Ga in GaN and InGaN quantum dots |
| Dark-field objective | Unique in range; detects particles below 100 nm |
| Confocal depth profiling | Stress distributions in SiC and GaN; SiC polytypes |
| Spectral resolution | Better than 1.2 cm⁻¹; 1.2 cm⁻¹ FWHM at 785 nm/1200 lines/mm; 0.4 cm⁻¹ per pixel |
| Peak-position accuracy | 0.1 cm⁻¹ (532 nm, 2400 lines/mm) |
| Spectral range | From 100 cm⁻¹; optional 50 cm⁻¹ |
| Spatial resolution | X 350 nm; Y 500 nm; Z 1 µm (532 nm, 100× objective, NA 0.90) |
| Lasers | 325, 355, 488, 532, 671, 785 nm and others |
| Spectrometer | 550 mm focal length |
| Gratings | 150, 300, 600, 1200, 1800, 2400 lines/mm; up to three installed |
| Detector | Thermoelectrically cooled 1340 × 400 CCD |
| Objectives | 5×, 10×, 20×, 50×, 100× and dark-field |
| Dimensions | 1276 × 1320 × 1350 mm; 390 kg including vibration-isolation table |