RAMANwalk
Chemical-information-driven beam routing gives a Raman preview 5–10 times faster than grid scanning.
RAMANwalk uses a random-walk algorithm for fast preview maps, 350 nm lateral resolution and a 1650-pixel detector.
RAMANwalk speeds confocal Raman mapping through an algorithm rather than new optics. Conventional grid scans spend equal time on empty and interesting areas; its random-walk method follows chemical spectra, steering the beam around detected components to build a preliminary image before a full map. Bruker claims a 5–10-fold preview speed gain.
Point illumination uses galvanometer X/Y scanning and stage Z motion. Spectral resolution is better than 1.5 cm⁻¹ depending on grating, with 1.5 cm⁻¹ FWHM at 785 nm/1200 lines/mm and 0.74 cm⁻¹ per pixel. Range begins at 100 cm⁻¹. With a 100×/0.90 NA objective, resolution is 350 nm in both X and Y and 1 µm Z.
One or two lasers can be chosen from 488, 532, 671 and 785 nm. A cooled 1650-pixel CCD line detector matches the point-scanning architecture. Polarization, low-wavenumber access from 50 cm⁻¹ and precision peak-shift tracking are options. Manufacturer materials differ on stage specification, so confirm the selected configuration.
| Manufacturer | Bruker Corporation (Nanophoton, a Bruker Company) |
|---|---|
| Type | Confocal Raman microscope with random-walk search |
| Illumination | Point |
| Scanning | Galvanometer X/Y; stage Z |
| Random-walk algorithm | Chemically guided sample search; beam route optimized around characteristic signals using stochastic-process and information theory |
| Speed gain | Manufacturer claims 5–10 times faster preview than conventional scanning Raman microscopy |
| Spectral resolution | Better than 1.5 cm⁻¹ depending on grating; 1.5 cm⁻¹ FWHM at 785 nm/1200 lines/mm; 0.74 cm⁻¹ per pixel |
| Spectral range | From 100 cm⁻¹ |
| Spatial resolution | X/Y 350 nm; Z 1 µm (100× objective, NA 0.90); better Y than RAMANtouch/RAMANdrive (350 vs 500 nm) |
| Lasers | Choice of one or two from 488, 532, 671 and 785 nm |
| Gratings | 300 and 1200 lines/mm; up to three installed |
| Detector | High-sensitivity thermoelectrically cooled 1650-pixel CCD line array |
| Objectives | 5×, 10×, 20×, 50×, 100× |
| Stage | Manufacturer sources differ: Bruker lists motorized XYZ 30 × 30 × 35 mm; Nanophoton lists manual XY and motorized Z, optional motorized XY |
| Options | Polarization; low-wavenumber range from 50 cm⁻¹; high-precision peak-shift tracking |
| Dimensions | 304 × 829 × 448 mm |