S6 JAGUAR
Wavelength-dispersive XRF resolution on a bench without an external chiller.
S6 JAGUAR covers F–Am using a 400 W tube, up to four analyzer crystals and 24-position EasyLoad.
S6 JAGUAR is a benchtop wavelength-dispersive XRF for industrial and research elemental work, designed for plug-and-analyze installation. WDXRF resolves lines better than EDXRF and supports light elements from fluorine through americium.
Up to four crystal positions can hold XS-55 for F–Mg, PET for Al–Cl, LiF200 for K–U or other crystals. An internal temperature-control unit eliminates an external water chiller, and a vacuum pump is included. SMART-QUANT WD provides fundamental-parameters standardless analysis; GEO-QUANT Basic serves geology. Uses include cement, food/feed, glass/ceramics, materials, metals, mining, petrochemicals and pharmaceuticals.
| Instrument type | Benchtop wavelength-dispersive XRF (WDXRF) |
|---|---|
| Element range | F–Am |
| X-ray tube | LongLife 400 W, up to 50 kV and 17 mA |
| Detectors | Proportional counter; HighSense XE (Ti–Am); 2 million counts/s linear range |
| Analyzer crystals | Up to four positions: XS-55 (F–Mg), PET (Al–Cl), LiF200 (K–U), Ge, XS-Ge-C, XS-400 or LiF220 |
| Analyzed spot diameter | 34 mm standard; optional 28, 23 or 18 mm |
| Primary-beam filters | Optional five-position changer |
| Measurement atmosphere | Vacuum (pump included, 265 × 180 × 489 mm); optional helium |
| Sample loading | Manual; 24-position EasyLoad; 22-position EasyLoad ONLINE; 15-position XY changer for two-part cells |
| Dimensions and weight | 51 × 67 × 74 cm (H×W×D), 132 kg manual; 80 × 67 × 74 cm, 147 kg EasyLoad/XY |
| Power | 100–240 V, 50/60 Hz, up to 1 kVA |
| Cooling | Integrated temperature control; no external chiller |
| Software | SMART-QUANT WD (standardless fundamental parameters), GEO-QUANT Basic; optional TouchControl |
| Safety | CE, DIN EN ISO 9001:2008; radiation below 1 µSv/h |