S8 TIGER Series 2
Floor-standing sequential WDXRF from Be to Am with three tube powers and 300 µm XRF mapping.
S8 TIGER Series 2 covers Be–Am, offers 1/3/4 kW tubes, up to eight crystals and XRF2 mapping at 300 µm.
S8 TIGER Series 2 is a floor-standing sequential wavelength-dispersive XRF for industrial and research elemental analysis. HighSense covers Be–Am for QC and process control.
The Rh or Cr tube comes in 1, 3 or 4 kW versions. The 1 kW model needs no cooling water; 3/4 kW versions do. Eight crystal positions and DynaMatch extend linear count range above 13 million counts/s.
XRF2 mapping uses a 300 µm spot and 100 µm step to examine sample heterogeneity instead of only bulk composition. Applications include cement, minerals, mining, metals, glass/ceramics, petrochemicals, polymers and research.
| Instrument type | Floor-standing sequential wavelength-dispersive XRF |
|---|---|
| Element range | Be–Am |
| X-ray tube | 1, 3 or 4 kW; Rh/Cr anode; 50 kV/50 mA (1 kW), 60 kV/150 mA (3 kW), 60 kV/170 mA (4 kW) |
| Analyzer crystals | Up to eight; XS-55, PET, LiF(200) standard; optional XS-Ge-C, XS-CEM, XS-400 and others |
| Detectors | Flow proportional counter; scintillation counter (>3 million counts/s, 2 million linear); DynaMatch linear range above 13 million counts/s |
| Analyzed spot diameter | Automatic masks: closed SampleCare, 34, 28, 23, 18, 8 and 5 mm; 300 µm probe mask |
| XRF2 mapping | 300 µm spot with 100 µm step; alternatively 1.25 mm |
| Changers | Four collimators; ten primary filters |
| Measurement atmosphere | Vacuum standard; atmospheric helium; optional vacuum seal |
| Detector gas | P5 (5% methane/95% argon) or P10 (10% methane/90% argon) |
| Sample loading | Automatic EasyLoad magazine |
| Dimensions | 135 cm H × 84 cm W × 90–104 cm D by power version |
| Weight | 446 kg (1 kW); 476 kg (3/4 kW) |
| Power | 1 kW: 208–240 V, single/three phase, 50/60 Hz; 3/4 kW: 208 V 60 Hz or 230 V 50/60 Hz three-phase |
| Cooling | 1 kW no water; 3/4 kW require automatically controlled cooling-water flow |
| Software | Multilingual TouchControl, QUANT-EXPRESS and XRF2 mapping |
| Safety | CE, DIN EN ISO 9001:2008; radiation below 1 µSv/h |