solariX
High-field FT-ICR workhorse with magnets up to 15 T for petroleum and other ultracomplex mixtures.
solariX offers 7, 12 and 15 T FT-ICR configurations, resolving power above 10 million and mass accuracy down to 250 ppb.
solariX is Bruker's high-field magnetic resonance mass spectrometry platform. Available with 7, 12 or 15 T magnets, it achieves resolving power above 10 million in every configuration. Mass accuracy improves from 600 ppb at 7 T to 250 ppb at 15 T.
Its principal application is ultracomplex mixtures containing tens of thousands of components, including petroleum and dissolved natural organic matter. MRMS, the ParaCell and magnetron-motion control enable routine access to isotopic fine structure across a broad mass range, separating ions whose isotopic composition differs despite the same nominal mass.
ESI is standard. Options include a dual ESI/MALDI source with a smartbeam-II laser and ETD fragmentation. Unlike scimaX, solariX requires annual cryogen refills and a quench line. For imaging, solariX XR is specified to exceed 200,000 resolving power at m/z 800 at a 1 Hz scan rate.
| Instrument type | Fourier transform ion cyclotron resonance mass spectrometer (FT-ICR/MRMS) |
|---|---|
| Magnetic field options | 7 T, 12 T, 15 T |
| Maximum resolving power | More than 10,000,000 in all configurations |
| Mass accuracy (internal calibration) | 600 ppb at 7 T; 300 ppb at 12 T; 250 ppb at 15 T |
| Detector | ParaCell XR (12 and 15 T); ParaCell XR or 2xR (7 T) |
| Ion sources | ESI standard; optional MALDI (dual ESI/MALDI source with smartbeam-II laser) |
| Fragmentation | Optional ETD |
| Cryogens | Annual refill required; quench line required |
| Isotopic fine structure (IFS) | Routine access over a broad mass range |
| Imaging (solariX XR) | Resolving power above 200,000 at m/z 800 and 1 Hz; CASI selected-ion accumulation option with up to 14 windows |
| Typical configurations | 7 T for MALDI imaging and flow-injection analysis; 15 T for petroleum analysis |