UniversalExtractor E-800
Four extraction methods in one unit, water or organic solvents, and >90% solvent recovery.
UniversalExtractor E-800 has six positions and an inert-gas Pro version.
UniversalExtractor E-800 covers hot, Soxhlet, continuous and Twisselmann extraction with optional chamber heating. Standard, Pro and Pro Inert versions extend methods and oxygen-sensitive sample protection. Efficient condensers recover over 90% of solvent.
| Manufacturer | BÜCHI Labortechnik AG, Switzerland |
|---|---|
| Methods | Hot extraction with/without chamber heat; conventional/warm Soxhlet with/without; continuous with/without; heated Twisselmann |
| Versions | Standard, Pro extended methods, Pro Inert with protective gas |
| Positions | Six |
| Solvents | Water and organic solvents boiling below 150 °C |
| Solvent recovery | >90% |
| Maximum chamber fill | Universal 190 mL; LSV 315 mL |
| Beaker volume | Universal 175 mL; LSV 320 mL; hot extraction 100 mL |
| Thimbles | 25 × 150 or 33 × 150 mm |
| Reservoir | 2 L |
| Power | 200–240 V AC ±10%, 50/60 Hz; 10 A fuse |
| Consumption | 1780 W |
| Cooling water | ≥100 mL/min at nominal 6 bar; inlet 25 °C below solvent boiling point |
| Operating conditions | 5–40 °C, humidity 80% to 31 °C tapering to 50% at 40 °C; altitude 2000 m |
| Glass deactivation | Heat at 450 °C |
| Dimensions/weight | Without glass 638 × 595 × 613 mm, 44.8 kg; with universal glass 638 × 595 × 752 mm, 52.6 kg |
| Method memory | 40 |
| Languages | German, English, Italian, Spanish, French, Japanese, Chinese, Polish, Russian |
| Methods | EPA 3541 HE; AWWA Standard Methods 5520 D |