ORACLE
Fat in 30 seconds by proton TD-NMR without solvent or product-specific calibration.
ORACLE can analyze up to 100 samples a day with optional robotic loading.
ORACLE detects hydrogen-proton signals from fat molecules by time-domain NMR. It avoids extraction solvents and product-specific method curves. The manufacturer gives differing lower limits across documents; those figures need confirmation for a particular application.
| Manufacturer | CEM Corporation, USA |
|---|---|
| Method | Time-domain NMR with CPMG pulse sequence |
| Range | To 100% fat; lower limit variously 0.01%, 0.02% or 0.05% in manufacturer documents |
| Product-specific calibration | Not required |
| Solvents | None |
| Time | 30 s/sample; full result <1 min |
| Throughput | To 100/day |
| Automation | Optional 100-sample robot; heating blocks for batches to 50 |
| Cost per test, manufacturer estimate | ~US$0.96 vs $12–14 Soxhlet or $11–13 Mojonnier |
| Methods | ISO 16756 / IDF 259; AOAC 2008.06; AOAC PVM 1:2004 |
| Reference correlation | Mojonnier, Weibull–Berntrop, Röse–Gottlieb |
| Independent validation | Actalia Cecalait |
| Paired workflow | With CEM SMART 6, moisture/solids/fat in under 5 min |
| Data export | USB and LIMS |