ANALYSETTE 22 NeXT Micro
Laser diffraction from 0.5 to 1500 µm with one reverse-Fourier scan in about 10 seconds.
ANALYSETTE 22 NeXT Micro supports wet and dry dispersion under ISO 13320.
ANALYSETTE 22 NeXT Micro measures particle-size distributions by static light scattering with a fixed reverse-Fourier optical layout. Its 51-element detector covers 0.5–1500 µm without switching lenses. Wet recirculation, ultrasound and a dry Venturi feeder support powders and suspensions.
| Method | Static light scattering / laser diffraction |
|---|---|
| Range | 0.5–1500 µm |
| Models | Fraunhofer and Mie |
| Standard | ISO 13320 |
| Optics | Reverse Fourier |
| Laser | 520 nm green solid-state diode, ~1 mW, 10,000 h life |
| Intensity stability | <2% |
| Beam | Coherent Gaussian; linear polarization >100:1 |
| IEC 60825-1 laser class | Class 1 |
| Laser alignment | Automatic |
| Lens | 400 mm focal length, 360 mm working distance; fixed optics |
| Detector | Custom 51-element semiconductor, half-ring/segment geometry |
| Wide-angle/backscatter | Absent |
| Overload level | 16-bit, 65,536 |
| Measurement time | Typically 5–10 s; full cycle ~1 min |
| Wet module | Closed 150–500 mL loop, radial pump to 3.5 L/min, 4 mm path; 10 mg to several grams sample |
| Wet-path materials | 316L stainless, PTFE, FFKM, FEP, BK7 glass, FKM and silicone tubing; optional chemical-resistance kit |
| Ultrasound | Adjustable to 50 W at 40 kHz |
| Dry module | Venturi nozzle, high-frequency feeder, 1–300 cm³ sample; clean dry oil-free air 4–8 bar/100 L/min; external extraction required |
| Cell | Open 7–8.5 mL, 0.5–250 µm range, magnetic stirrer |
| Additional records | Temperature and pH during analysis; configurable SOPs |
| Software | NeXT Control; PC, screen, keyboard and mouse included |
| PC | Windows 10+, 4 GB RAM, USB |
| Power | 100–240 V, 50–60 Hz, 120 W |
| Measurement unit | 670 × 350 × 320 mm W×D×H; 19.2 kg |
| Order number | 22.9000.00 |