IB / IL incubators
IB heats from ambient; refrigerated IL cools to 0 °C.
JEIO TECH IB4/IL3 incubators span 25–254 L; IB operates to 80 °C and IL supports cooled culture and BOD tests.
Forced convection or air-jacket configurations provide options for sample protection and temperature stability.
| Instrument type | Laboratory incubators; IB heating only, IL compressor cooling |
|---|---|
| IB4 forced-convection models | IB4-03/-05/-10/-15 in S and V versions |
| IB4-A air-jacket models | IB4-03A/-05A/-10A/-15A in AS and AV versions |
| IL3 models | IL3-15/-25 forced convection and IL3-15A/-25A air jacket; ILP-02/-12 personal 48 L |
| IB4 temperature | Ambient +5 to 80 °C |
| IL3 temperature | 0–80 °C; air-jacket versions 4–70 °C |
| Fluctuation at 37 °C | ±0.1–0.15 °C IB4; ±0.1–0.2 °C IL3 |
| Uniformity at 37 °C | ±0.23–0.55 °C IB4; ±0.7–1.0 °C IL3 |
| Heat-up to 37 °C | 4–8 min forced IB4; 40–49 min air-jacket IB4-A |
| Chamber capacity | 25–150 L IB4; 150–254 L IL3 |
| IB4 internal dimensions (W × D × H) | 300 × 210 × 400 to 550 × 376 × 710 mm |
| IL3 internal dimensions (W × D × H) | 600 × 500 × 500 to 500 × 540 × 940 mm |
| Shelves | IB4: two standard, up to 4–9, 20 kg per shelf; IL3: two/three standard, up to 6–25, 26 kg on IL3-15 |
| Weight | 31–62 kg IB4; 100–145 kg IL3 |
| IB4 power | 230 V 50/60 Hz at 2.5–3.5 A; 120 V 60 Hz at 4.8–6.7 A |
| IL3 power | 230 V 50 Hz at 4.5–8.0 A; 120 V 60 Hz at 8.3–12.3 A |
| Controller | IB4: 3.5/5-inch TFT; IL3: PID/VFD with three temperature memories; self-tuning PID, three-point calibration and Wait On/Off timer |
| Event log | 12 S/AS or 36 V/AV events |
| Interfaces | USB-B, RS-232 standard on V/AV and optional on S/AS; IL3 USB/RS-232 with control software |
| Protection | Patented overheat prevention, independent limiter, high/low alarms, open-door alarm and restart within 30 s of power restoration |
| IL3 construction | Stainless chamber, tempered-glass inner door with silicone seal, internal power outlet |
| Standards | DIN 12880 Class 3.1, IEC Class II; optional LC DataKeeper for 21 CFR Part 11 |
| Options | Water tray, two-tier stacking, recorder, IL3 CO2 control and LC GreenBox |