Plant growth chambers
Control lighting to 35,000 lux and temperature to ±0.1 °C.
JEIO TECH GC-300/GC-1000TLH: 300 or 1000 L, 5–50 °C, optional humidity and CO2 control to 5000 ppm.
Side and top lighting plus heat exhaust support repeatable plant growth conditions.
| Instrument type | Plant growth environmental chambers |
|---|---|
| Models | GC-300TL no humidity, GC-300TLH, GC-1000TLH; T temperature, L light, H humidity |
| Chamber volume | 300 L GC-300; 1000 L GC-1000TLH |
| Controls | Microprocessor PID |
| Temperature | 5–50 °C lights off; 10–50 °C lights on; 20–50 °C with humidity control |
| Fluctuation at 25 °C | ±0.1 °C without humidity |
| Uniformity at 25 °C | ±0.8 °C without humidity |
| Illuminance | 0–20,000 lux GC-300; 0–35,000 lux GC-1000TLH |
| Light settings | Ten steps GC-300; eleven GC-1000TLH |
| Lamps | GC-300: 18 × 32 W fluorescent; GC-1000TLH: 16 × 32 W fluorescent plus 3 × 400 W metal halide |
| Relative humidity | None on 300TL; 300TLH: 40–80% at 20–35 °C or 70–90% at 36–50 °C; 1000TLH: 50–90% or 60–90% |
| Humidity fluctuation at 60% RH | ±3% RH |
| Optional CO2 control | To 5000 ppm with NDIR sensor |
| Refrigeration | 1/2 hp GC-300; 3/4 hp GC-1000TLH |
| Temperature heater | 2 × 750 W GC-300; 2 × 1500 W GC-1000TLH |
| Humidification heater | 1300 W GC-300TLH; 1500 W GC-1000TLH |
| Internal dimensions (W × D × H) | 510 × 540 × 1100 mm GC-300; 1200 × 800 × 1080 mm GC-1000TLH |
| External dimensions (W × D × H) | 700 × 805 × 1900 mm GC-300; 1410 × 1070 × 2150 mm GC-1000TLH |
| Net weight | 250 kg GC-300; 550 kg GC-1000TLH |
| Power | 230 V; 300TL 12.5 A, 300TLH 16 A, 1000TLH 30 A at 60 Hz or 23.2 A at 50 Hz |
| Wire shelves | GC-300: three of 14 at 466 × 490 mm; GC-1000TLH: six of 29 at 580 × 770 mm; perforated shelves available |
| Programming | Ten temperature/humidity/light steps repeated up to 999 cycles; temperature calibration and autotuning |
| Construction | Sealed inner glass and outer magnetic doors, refillable water tank, side/top lights, top heat exhaust and high-frequency ballasts |
| Options | Recorder and CO2 sensor |