TC3 heating and cooling chambers
Benchtop temperature cycling from −40 to 100 °C with ±0.3 °C fluctuation.
JEIO TECH TC3: 25, 65 or 100 L chambers without humidity control, offering up to 100 patterns and 2000 segments.
The TC3 platform shares controls with TH3 temperature/humidity chambers for temperature-only testing.
| Instrument type | Benchtop heating/cooling test chambers without humidity control |
|---|---|
| Models | TC3-KE-025/065/100 and TC3-ME-025/065/100 |
| TC3-KE temperature | −35 to 100 °C at 50 Hz; −40 to 100 °C at 60 Hz |
| TC3-ME temperature | −20 to 100 °C |
| Temperature fluctuation | ±0.3 °C |
| Temperature uniformity | ±0.5 °C |
| Heating time | 60 min KE; 50 min ME |
| Cooling time | 70 min KE; 55 min ME |
| Humidity control | None |
| Chamber volume | 25, 65 or 100 L |
| Internal dimensions (W × D × H) | 350 × 220 × 350; 400 × 330 × 495; 500 × 380 × 530 mm |
| External dimensions (W × D × H) | 550 × 880 × 915; 600 × 1050 × 1060; 700 × 1110 × 1100 mm |
| Shelves | Two standard, up to 7 or 10; 20 mm spacing; 25 kg per shelf and 65 kg total |
| Cable port | Ø50 mm |
| Chamber material | Stainless steel |
| Weight | KE 110/140/150 kg; ME 80/110/130 kg |
| TC3-KE power | 230 V single phase 50/60 Hz at 8.5/9.5/10.6 A |
| TC3-ME power | 230 V single phase 50/60 Hz at 5.5/6.5/8.6 A |
| Controller | 5.6-inch color touchscreen |
| Programming | 100 patterns, 999 repeats, 255 partial repeats, 100 segments/pattern, 2000 total, 99 h 59 min/segment |
| Interfaces | RS-232 standard, RS-485 optional; up to 32 instruments/PC |
| Standards | Optional LC DataKeeper for 21 CFR Part 11 GMP audit |
| Options | One/two-level stand, paper/digital recorder, LC Connected/GreenBox mobile monitoring |