TH / TH3 temperature and humidity chambers
Program temperature down to −70 °C and humidity from 20–95% RH.
JEIO TECH TH/TH3 chambers span 25–1450 L, with up to 100 patterns and 2000 segments on TH3.
Benchtop and floor-standing versions support cyclic environmental testing with controlled temperature and humidity.
| Instrument type | Environmental temperature and humidity test chambers |
|---|---|
| Series and models | Benchtop TH3-KE/ME/PE 025/065/100; TH3-E 200/400; TH3-G 180/408; TH3-TG 180/408/800/1000/1500; TH-G 800/1000/1500; TH-KH 180/300/408 |
| TH3-KE temperature | −35 to 100 °C at 50 Hz; −40 to 100 °C at 60 Hz |
| TH3-ME/PE/E temperature | TH3-ME −20 to 100 °C; PE 0–100 °C; E 0–90 °C |
| TH3-G/TG temperature | G −40 to 100 °C; TG −5 to 100 °C |
| TH-G/KH temperature | TH-G −40 to 100 °C at 60 Hz; TH-KH −70 to 100 °C dry or 15–90 °C with humidity |
| Temperature fluctuation | ±0.3 °C benchtop TH3/TH; ±0.5 °C TH3-E/G/TG |
| Temperature uniformity | ±0.5 °C benchtop; ±0.8 °C TH3-E; gradient <2.0 °C TH3-G/TG |
| Heating/cooling rate | 1.0 °C/min TH3-G; 0.5 °C/min TH3-TG; other models specify times instead |
| Heat-up 0 to 90 °C | 40–60 min benchtop TH3; 80 min TH3-E |
| Relative humidity range | 20–95% RH KE/ME/TG; 30–95% PE; 35–85% E; 25–95% G/TH-G/TH-KH |
| Humidity fluctuation | ±1% RH benchtop/E; ±2.5% G/TG; ±0.4% KH |
| Humidity uniformity | ±3% RH |
| Chamber volume | 25/65/100 L benchtop; 180/189/201/302/398/408/800/1000/1450 L floor-standing |
| Internal dimensions | 350 × 220 × 350 to 1000 × 1000 × 1450 mm |
| Shelves | Two standard, up to 7–19; 20–40 kg per shelf, total 50–100 kg |
| Water tank | 8–20 L; water pH 6.2–7.2, conductivity ≤20 µS/cm |
| Weight | 80–860 kg depending on model |
| Power | 230 V single phase 7.2–30 A; large models 380 V three phase 10.1–15.5 A |
| Controller | 5.6-inch color touchscreen KE/ME/PE/E; 5.7-inch TH-G/KH |
| TH3 programming | 100 patterns, 999 repeats, 255 partial repeats, 100 segments/pattern, 2000 total, 99 h 59 min/segment |
| TH programming | 120 patterns, 999 repeats, 255 partial repeats, 100 segments/pattern, 1200 total, 999 h 59 min/segment |
| PID control | Nine zones TH3-E; 16 zones TH3-G/TG |
| Interfaces | RS-232 standard, RS-485 optional; up to 32 instruments per PC |
| TH refrigeration | Two-stage air-cooled compression; R-507, plus R-23 on TH-KH |
| Cable port | Standard Ø50 mm; up to two optional Ø50/80 mm ports |
| Safety | Electronic/mechanical overheat protection, surface ≤60 °C, low-water/open-door alarms, overload/short-circuit and refrigerant-pressure protection |
| Standards | ICH Q1A(R2) for TH3-TG; optional LC DataKeeper 21 CFR Part 11 integration |
| Options | Recorder, inner glass door/hand-in port, direct water feed, IQ/OQ/PQ and LC Connected/GreenBox monitoring |
| Other features | Heated tempered door glass with LED light, locking handle, temperature/humidity autotuning and external humidifier |