BW water and oil baths
Water baths of 3.5–20 L; stirred versions hold ±0.1 °C at 50 °C.
JEIO TECH BW heating baths reach 100 °C, while VB oil baths reach 150 °C.
General, stirred, economy and shaking configurations cover routine thermostating.
| Instrument type | Laboratory heating baths |
|---|---|
| Versions | BW3-G General 05/10/20G; BW-H stirred 05/10/20H; BW-B Economy 05/20B; VB Visible 25/40G to 150 °C; BS2 shaking 10/15/20/30 |
| BW3-G temperature | Ambient +7 to 100 °C |
| BW-H/BS2 temperature | Ambient +5 to 100 °C |
| VB temperature | Ambient +5 to 150 °C |
| Stability at 50 °C | ±0.3–0.4 °C BW3-G; ±0.1 °C BW-H/VB/BS2 |
| Heat-up to 70 °C | 34–44 min BW3-G; 25–50 min BW-H; 35–55 min VB |
| Bath volume | BW 3.5 / 11.5 / 20 L; VB 25/40 L; BS2 17/26/37/55 L |
| Bath opening and depth | BW 240 × 136 × 150, 300 × 240 × 200 or 498 × 300 × 200 mm; VB 145 × 265 or 295 × 265 × 330 mm |
| BS2 shaking speed | 30–200 rpm; 25 mm stroke on 10/15 or 30 mm on 20/30 |
| Bath material | 1.0 mm stainless steel; double-coated, baked-steel housing |
| Temperature sensor | Pt100 |
| Controller | Microprocessor PID, temperature calibration and timer |
| Protection | Overheat protection and limiter, low-liquid alarm, mechanical thermostat |
| Weight | BW3-G 6.5/10.0/18.5 kg; BW-H 6.5/9.5/15.5; VB 28/37; BS2 23–38 |
| Power | 230 V 50/60 Hz at 3.0–8.7 A BW or 13 A VB; 120 V 60 Hz at 5.8–8.3 A |
| Note | BW3-05G is supplied without a drain valve |
| Circulation pump | Not fitted to BW; circulation is available in CW3/RW3 series |