BELCAT II (chemisorption, TPD/TPR/TPO)
Automated catalyst characterization with furnace to 1200 °C and rapid cooling.
BELCAT II combines TPD, TPR, TPO, pulse chemisorption and BET.
BELCAT II is a flow-through catalyst analyzer for temperature-programmed desorption, reduction and oxidation. Pulse chemisorption and breakthrough curves add active-site and gas-interaction information. The furnace cools from 400 to 50 °C in 30 minutes; an optional mass spectrometer follows evolved species.
| Instrument type | Fully automated catalyst chemisorption analyzer |
|---|---|
| Methods | TPD, TPR, TPO, pulse chemisorption, breakthrough curves and BET |
| Principle | Dynamic flow |
| Detector | High-sensitivity four-element semidiffusion TCD |
| Measurement/pretreatment ports | One; triple-cylindrical cell with gas preheat |
| Dose gases | He, Ar, N₂, O₂, H₂, CO, CO₂, NH₃, N₂O, NO and others |
| Carrier lines | Three |
| Pretreatment/pulse/MIX lines | Eight, two corrosion resistant |
| Flow controllers | 100 sccm full scale carrier and pretreatment/pulse; 30 sccm MIX |
| Furnace | Normally to 1100 °C, maximum 1200 °C |
| Rapid cooling | 400→50 °C in 30 min |
| CATCryo II option | To −120 °C |
| Vapor option | Water, methanol, ethanol, toluene, benzene and others |
| Gas pressure | Measurement 0.1 MPa gauge; valve drive 0.45–0.55 MPa gauge |
| Connection | Swagelok 1/8 inch |
| Dimensions (W×H×D) | 280 × 400 × 600 mm; 36 kg |
| Power | 110 or 220 V AC, 1300 W |
| BELMASS II option | Online MS 1–200 amu, Faraday/electron multiplier, gas-dependent <1 ppm detection, M/ΔM ≥2M, 16 channels, RS-232C, sample 50–150 kPa at ~0.6 cm³/min |