BELSORP MINI X
Four parallel sorption ports with independent pressure sensors measure surface area from 0.01 m²/g.
BELSORP MINI X analyzes gas physisorption and pore sizes from 0.7 to 500 nm.
BELSORP MINI X uses volumetric gas sorption with continuous AFSM free-space measurement. Four sample ports can operate together, with separate pressure sensors. Quick BET serves QC, while high-precision mode and gas choices support research.
| Instrument type | Gas-physisorption surface-area and pore analyzer |
|---|---|
| Principle | Volumetric manometric with AFSM free-space measurement |
| Measurement ports | Four parallel; three-port version available |
| Pressure sensors | Six, or five for three-port; 133.3 kPa/1000 Torr; one per sample plus saturation-pressure port |
| Gas ports | Two standard, up to five optional |
| Surface area | From 0.01 m²/g N₂; 0.0005 m²/g Kr |
| Pores | 0.7–500 nm; optional 0.35 nm molecular probe |
| Relative pressure P/P₀ | 1 × 10⁻⁴–0.997 for N₂ at 77 K or Ar at 87 K |
| Adsorbates | N₂, Ar, Kr, CO₂, H₂, CH₄, butane and other noncorrosive gases |
| Vapor sorption | Unsupported |
| Sample temperature | LN₂ or liquid Ar; 2 L Dewar holding 30 h; water bath −10–70 °C (English brochure says −10–80 with circulator) |
| Pretreatment temperature | 50–450 °C |
| Modes | High-precision research and Quick BET QC |
| Preparation stations | BELPREP VAC II three ports to 430 °C or VAC III six to 450 °C |
| Standards | ISO 9277 BET, ISO 15901-2 pores, JIS Z8830/Z8831-2; optional 21 CFR Part 11 for selected models |
| Gas supply | He/adsorbate 0.1 ±0.02 MPa |
| Dimensions (W×H×D) | 280 × 650 × 465 mm |
| Weight | 31 kg main unit; 38 kg series catalog, manufacturer discrepancy |
| Power | 100–240 V AC, ≤700 W excluding rotary pump; series catalog says 1000 W |
| Certification | CE |