BELSORP MR 1
Single-point BET in about 15 minutes without a vacuum system.
BELSORP MR 1 measures routine surface area from 0.01 m²/g by flowing gas.
BELSORP MR 1 is a rapid flowing-gas BET analyzer for QC. It uses nitrogen or krypton adsorption with helium carrier and thermal-conductivity detection. It determines surface area but does not yield full pore-size distributions or adsorption isotherms.
| Manufacturer | Microtrac Retsch GmbH (MICROTRAC MRB), Germany |
|---|---|
| Type | Fast BET surface-area analyzer |
| Replaces | BELSORP MR6; former address now displays MR 1 |
| Method | Dynamic flowing-gas single-point BET with TCD |
| Surface area | From 0.01 m²/g |
| Repeatability | Within ±1.0% |
| Stations | One |
| Adsorbate/carrier | Nitrogen or krypton; helium carrier |
| Degassing | To 400 °C |
| Measurement time | ~15 min including calibration, excluding pretreatment |
| Vacuum | Not required |
| Method limits | No pore distribution or full isotherm; use MINI X/MAX X |
| Dimensions | 350 × 553 × 368 mm; 30 kg |
| Power | 100–240 V AC, 400 W, 50/60 Hz |