SYNC (laser diffraction and imaging)
Particle size by laser diffraction and shape by dynamic imaging in a single run.
SYNC GO/PLUS/PRO measures down to 0.1/0.02/0.01 µm and records over 30 shape metrics.
SYNC combines Tri-Laser diffraction and a 5.2 MP camera to characterize the same sample in one analysis. Wet and dry modules interchange quickly; DIMENSIONS LS reports size distributions and morphology. The GO, PLUS and PRO versions differ in lower detection limit.
| Method | Laser diffraction plus dynamic image analysis in one measurement |
|---|---|
| SYNC GO range | 0.1–2000 µm |
| SYNC PLUS range | 0.02–4000 µm |
| SYNC PRO range | 0.01–4000 µm |
| Optics | Tri-Laser, detection to 165° |
| Red laser | 780 nm, nominal 0.35–2 mW |
| Blue laser | 405 nm, nominal 4–8 mW |
| Detectors | Two fixed photoelectric detectors with 151 logarithmic segments, 0–165°; brochure gives 0.02–165° |
| Camera | 5.2 MP 2560 × 2048, to 60 frames/s |
| Shape | More than 30 dimensional/morphological metrics per particle |
| Wet mode | Nominal 200 mL, flow 0–65 mL/s |
| Dry mode | To 100 psi/689 kPa; TURBOSYNC brochure says 50 psi, manufacturer discrepancy |
| Module exchange | <15 s |
| Measurement time | ~10–30 s |
| Standards | ISO 13320, 13322-2, 9276-6; FDA 21 CFR Part 11 |
| Software | DIMENSIONS LS |
| Operating temperature | 5–40 °C; dry storage −10–50 °C |
| Humidity | ≤90% noncondensing |
| Dimensions | ~820 × 460 × 500 mm; brochure says 457 × 787 × 305 mm, manufacturer discrepancy |
| Weight | FlowSync 19.5 kg, TurboSync 13.6, Sync 23.6; brochure says 24 kg |
| Power | 90–264 V AC, 47–63 Hz, single phase; 25 W nominal, 50 W maximum |