L / LT (muffle furnaces to 1200 °C)
Eight muffle-furnace sizes from 1 to 60 L, to 1200 °C and ±5 K uniformity.
Nabertherm L/LT uses replaceable ceramic heater plates and programmable B510 control.
L and LT are laboratory muffle furnaces with swing or lift doors. The 1–60 L range supports ashing and heat treatment to 1100 or 1200 °C. A double-wall ventilated body, protected heaters and B510 program controller support long-duration use.
| Instrument type | Laboratory muffle furnaces; L swing door, LT lift door |
|---|---|
| Maximum temperature | 1100 °C /11; 1200 °C /12 |
| Continuous-use recommendation | 1000 °C /11; 1100 °C /12 |
| Volumes | 1/3/5/9/15/24/40/60 L |
| Internal dimensions (W×D×H) | 90 × 115 × 110 mm 1 L to 380 × 490 × 330 mm 60 L |
| External dimensions (W×D×H) | 290 × 280 × 410 mm 1 L to 610 × 705 × 660 mm plus 385 mm door opening at 60 L |
| Connected power | 1.3–9.8 kW by size |
| Supply | Single-phase 1–15 L; three-phase 24/40/60; 110–240 V one phase or 200–240/380–480 V three, 50/60 Hz |
| Weight | 15/21/27/35/43/52/70/83 kg for increasing volumes |
| Time to Tmax−100 K | 25–160 min by size/version |
| Uniformity | ±5 K empty chamber to DIN 17052-1 above 800 °C |
| Heaters | Replaceable protected ceramic plates, two-sided or three-sided on large units |
| Thermocouple | Type N at 1100 °C; S at 1200 |
| Insulation | No EU Regulation 1272/2008 classified material; no RCF |
| Housing | Double-wall ventilated stainless, low outer temperature |
| Controller | B510 touch, five programs of four segments; R8 on 1 L |
| Switching | Solid-state relays |
| Interfaces | USB NTLog Basic, Wi-Fi MyNabertherm |
| Ventilation | Adjustable door intake/rear exhaust; optional chimney, fan or catalytic afterburner, NW80–120 |
| Options | Overtemperature limiter, protective gas, gas control, thermocouple port, VCD software, loading shelves to 800 °C |