LV / LVT (ashing furnaces)
Ashing furnaces exchange chamber air more than six times a minute.
LV/LVT preheat intake air behind the heaters and reach 1100 °C.
LV/LVT are dedicated ashing furnaces with forced air exchange to maintain oxygen during combustion. Intake air passes the heating zone before reaching samples; an integrated chimney removes exhaust. The family spans 3–15 L and uses B510 programmed control.
| Instrument type | Ashing furnaces; LV swing, LVT lift door |
|---|---|
| Maximum temperature | 1100 °C; ashing to 1050 °C |
| Volumes | 3/5/9/15 L: LV/LVT 3/11, 5/11, 9/11, 15/11 |
| Air exchange | >6 chamber volumes/min |
| Air path | Intake preheated behind heaters; integrated Ø80 mm exhaust |
| Maximum organic load | 5/10/15/25 g |
| Maximum evaporation rate | 0.1/0.2/0.3/0.3 g/min |
| Internal dimensions (W×D×H) | 180 × 150 × 120; 205 × 170 × 130; 235 × 240 × 170; 230 × 340 × 170 mm |
| External with chimney (W×D×H) | 345 × 390 × 810; 385 × 415 × 810; 415 × 485 × 865; 415 × 590 × 865 mm |
| Connected power | 1.3/2.6/3.3/3.6 kW |
| Supply | Single-phase 230 V |
| Weight | 20/29/36/44 kg |
| Time to Tmax−100 K | 45/55/70/80 min |
| Uniformity | To ±10 °C to DIN 17052-1 from 550 °C |
| Heaters | Replaceable protected ceramic plates, two-sided |
| Insulation | No EU 1272/2008 classified materials |
| Housing | Fan-cooled double-wall stainless, lower surface temperature |
| Controller | B510 touch, five four-segment programs |
| Interface | USB NTLog Basic |
| Standards | Suitable for ISO/ASTM/EN/DIN ashing; uniformity to DIN 17052-1 |
| Options | Overtemperature limiter, thermocouple port, shelves to 800 °C, VCD software |
| Related model | L .../11 BO with built-in exhaust treatment, 9/24/40 L, organic load 75/150/200 g, three-sided heat, C550 |