SW furnaces with integrated scales
Weigh samples inside the hot chamber while logging mass loss and temperature together.
NABERTHERM L 9/11/SW and L 9/12/SW: 9 L chamber, 1100 or 1200 °C and continuous in-chamber weighing with VCD software.
Integrated weighing removes the need to cool samples between measurements. External balances support capacities from 2200 to 12,000 g.
| Instrument type | Muffle furnace with integrated weighing for loss-on-ignition measurements |
|---|---|
| Models | L 9/11/SW, LT 9/11/SW, L 9/12/SW, LT 9/12/SW |
| Maximum temperature | 1100 °C (/11) or 1200 °C (/12) |
| Chamber volume | 9 L |
| Internal dimensions (W × D × H) | 230 × 240 × 170 mm |
| External dimensions (W × D × H) | 415 × 455 × 740 mm; plus 240 mm when the door is open |
| Connected load | 3.3 kW |
| Power supply | Single phase, 230 V |
| Weight | 50 kg |
| Heating time to 100 K below Tmax | 68 min (/11); 80 min (/12) |
| Weighing assembly | Ceramic rod and support plate built into the lining and supported by an external underbench balance; rod weight 850 g |
| EW-2200 balance | 0.01 g readability; 2200 g capacity; 0.1 g verification interval; 0.5 g minimum load |
| EW-4200 balance | 0.01 g readability; 4200 g capacity; 0.1 g verification interval; 0.5 g minimum load |
| EW-6200 balance | 0.01 g readability; 6200 g capacity; 1.0 g minimum load |
| EW-12000 balance | 0.10 g readability; 12,000 g capacity; 1.0 g verification interval; 5.0 g minimum load |
| Controller | B510; five programs with four segments each |
| Software | VCD for process control and recording and documenting temperature and mass loss |
| Alternative version | Muffle furnace with loss-on-ignition software but without a balance |