ARC 244
Simulate cooling failure and follow self-heating under adiabatic conditions.
ARC 244 measures reaction runaway to 500 °C with 20 K/min maximum tracking.
ARC 244 is an accelerating-rate calorimeter for thermal hazard assessment. Heat-Wait-Search identifies exothermic onset and then follows self-heating without removing reaction heat, simulating loss of cooling. Optional VariPhi adjusts thermal inertia; stirring, injection and pressure release broaden process experiments.
| Instrument type | Accelerating-rate adiabatic calorimeter |
|---|---|
| Temperature | Room temperature–500 °C |
| Pressure | 0–150 bar brochure; 0–200 bar manufacturer datasheet |
| Temperature reproducibility | 0.1 K |
| Maximum tracking rate | 20 K/min |
| Typical sample volume | 0.5–8.5 mL brochure; 0.5–7 mL datasheet; typical, not absolute limits |
| Cells | Spherical Hastelloy/stainless/titanium, 0.4–0.9 mm walls, 1–130 mL; tubular stainless/titanium, 0.4–0.9 mm, 1–8.5 mL |
| Modes | Heat-Wait-Search, Iso-Fixed/Iso-Track aging, linear screening, scanning/isothermal with power compensation, constant heating rate |
| Lift | Motorized |
| Thermal inertia | Optional VariPhi for low pressure-dependent phi factor |
| Options | Stirring, injection, pressure release, kinetic-analysis software |