ARC 305
Track fast exothermic reactions at up to 200 K/min with cells to 130 mL.
ARC 305 supports chemical-process, energetic-material and battery safety testing.
ARC 305 extends adiabatic runaway analysis to a 200 K/min tracking rate and larger sample cells. It records temperature and pressure during Heat-Wait-Search and other modes, including fire exposure. VariPhi can reduce the cell's thermal-inertia influence. Cells use materials suited to different chemistries.
| Instrument type | Accelerating-rate adiabatic calorimeter |
|---|---|
| Temperature | Room temperature–500 °C |
| Pressure | 0–150 bar brochure; 0–200 bar datasheet; 0.35% full-scale accuracy at 150 bar |
| Temperature reproducibility | 0.1 K |
| Maximum tracking rate | 200 K/min |
| Typical sample volume | 0.5–130 mL brochure; 1–130 mL datasheet |
| Cells | Spherical 1–130 mL; tubular 1–9 mL; stainless, Hastelloy, glass, titanium, tantalum; 0.4–0.9 mm walls |
| Modes | Heat-Wait-Search, constant heating, isothermal, adiabatic, scanning and fire-exposure |
| Lift | Motorized |
| Thermal inertia | Patented optional VariPhi for low pressure-dependent phi factor |
| Options | Stirring, injection, pressure release and kinetic software |
| Software | Proteus |
| Standard | Based on ASTM E1981 |
| Applications | Chemical processes, energetic materials and batteries |