DIL 502 Expedis (Supreme / Select / Classic)
Horizontal dilatometry from −180 to 2800 °C with 0.1 nm length-change resolution.
DIL 502 Expedis covers expansion of polymers, metals, ceramics and refractories across three versions.
DIL 502 Expedis measures thermally driven length changes with interchangeable furnaces. Supreme reaches 2800 °C and 0.1 nm resolution; Select reaches 2000 °C, while Classic covers routine work to 1600 °C. Different sample holders provide compression or tension and controlled atmospheres.
| Instrument type | Horizontal dilatometer |
|---|---|
| Versions | Classic, Select, Supreme |
| Temperature | Classic room temperature–1600 °C; Select −180–2000; Supreme −180–2800 |
| Length-change range | ±5 mm Classic; ±10 Select; ±25 Supreme |
| Length resolution | 2 nm Classic; 1 nm Select; 0.1 nm Supreme |
| Relative expansion reproducibility | 0.002% Classic; 0.001% Select/Supreme |
| Relative expansion accuracy | 0.003% Classic; 0.002% Select/Supreme |
| Applied force | 10 mN–3 N compression/tension by holder; 0.001 mN resolution |
| Sample | Length to 52 mm, 25 mm graphite furnace; standard Ø12 mm, optional Ø19; Ø8 dual holder |
| Heating rate | 0.001–50 K/min; Supreme graphite furnace to 100 K/min |
| Furnaces | Up to nine exchangeable: copper, stainless, quartz, SiC, Rh, graphite to 2800 °C |
| Atmospheres | Inert, oxidizing, reducing; vacuum Select/Supreme |
| Gas control | Single-channel flow, optional three/four-channel |
| Standards | DIN EN 821, DIN 51045, ASTM E228 |
| Materials | Ceramics, metals, composites, polymers, refractories |