GHP 456 Titan
Absolute guarded-hot-plate conductivity without reference standards, from −160 to 600 °C.
GHP 456 Titan covers 0.003–2 W/(m·K) for cryogenic and furnace insulation.
GHP 456 Titan measures insulation conductivity by the absolute guarded-hot-plate method, so no conductivity calibration standard is needed. Low- and high-temperature versions reach −160 to 250 or 600 °C. The symmetrical setup uses two samples, with a one-sample mode available; controlled vacuum and gas atmospheres enable pressure-dependent work.
| Instrument type | Absolute guarded hot plate (GHP) |
|---|---|
| Mean sample temperature | −160–250 °C low-temp; −160–600 °C high-temp |
| Conductivity | 0.003–2 W/(m·K) |
| Accuracy | Typically 2%; reproducibility <1% |
| Sample thickness | Up to 100 mm, typically 10–50 mm |
| Plate size | 300 × 300 mm |
| Minimum thermal resistance | 0.02 m²·K/W |
| Positions | Two identical samples symmetric; one-sample mode possible |
| Atmospheres | Oxidizing to 300 °C; inert, vacuum and controlled pressure |
| Vacuum | 5 × 10⁻⁴ mbar (0.05 Pa); controlled 0.1–100 mbar |
| Standards | ISO 8302, ASTM C177, DIN EN 12667/12939 |
| Materials | Cryogenic/furnace, fibrous/foam insulation, powders and flakes |