KBT 916 (cables)
Vertical cable-bundle fire test for European cable classification.
KBT 916 measures flame spread, heat release and smoke to EN 50399 and IEC 60332-3-10.
KBT 916 burns a vertical bundle of cables on a stainless ladder inside a full-height chamber. A propane ribbon burner provides a 20-minute exposure while the exhaust duct measures oxygen consumption, smoke and temperature. Software calculates heat and smoke growth indices for EN 50399 classification.
| Equipment type | Cable flame-propagation fire tester |
|---|---|
| Method | EN 50399 and IEC 60332-3-10 |
| Chamber | Double-wall welded steel with stainless facing/65 mm mineral wool, air/smoke openings, rails, electric winch and fire-glazed door |
| Chamber dimensions | 1350 × 2350 × 4300 mm; doorway 2160 × 1340 mm |
| Sample holder | Welded stainless tubular ladder 3600 × 500 mm |
| Burner | Venturi propane ribbon, piezo ignition/flame sensor, pivoting stainless cover; 7.5 cm from samples; 20 min |
| Exhaust hood | Stainless 1000 × 1500 × 1000 mm |
| Ventilation | Adjustable exhaust to constant flow; intake/exhaust fans with variable-frequency drives |
| Measurement duct | Heat-resistant stainless 1.4571 Ø400 mm; bidirectional flow probe, thermocouples, light attenuation and gas sampler |
| Gas analyzer | Siemens ULTRAMAT/OXYMAT 6E O₂/CO₂, optional CO; automatic calibration |
| Outputs | Heat-release rate/total, smoke-production rate/total, fire growth-rate index and duct temperature |
| Temperature range | −100–1300 °C, 24-bit ADC |
| Controller | Windows 10 KBT Control, 10.1-inch 1280 × 800 touchscreen, 16 configurable windows, 300 s pre-detection recording, automatic valves/burner |
| Weight | ~1900 kg apparatus; 168 kg main control cabinet; 110 kg console |
| Power | 230 V, 50 Hz |
| Standards | EN 50399, IEC 60332-3-10 |