LFA 707 StratoFlash
Laser-flash diffusivity up to 1600 °C with 2.5% stated accuracy.
LFA 707 StratoFlash measures 0.01–2000 mm²/s with a tunable 1054 nm laser pulse.
LFA 707 StratoFlash replaces the xenon flash with a pulsed neodymium-glass laser. It measures thermal diffusivity from room temperature to 1600 °C on round or square samples. The automated changer holds three or five samples, and a high vacuum or inert atmosphere is available.
| Instrument type | Laser-flash thermal diffusivity/conductivity analyzer |
|---|---|
| Temperature | Room temperature–1600 °C |
| Heating rates | 0.01–50 K/min |
| Diffusivity | 0.01–2000 mm²/s |
| Conductivity | 0.1–4000 W/(m·K) |
| Accuracy | ±2.5% diffusivity; ±5% specific heat |
| Repeatability | ±1% diffusivity; ±3% specific heat |
| Pulse source | 1054 nm pulsed Nd:glass laser, adjustable <0.05–1.5 ms and 0.25–25 J/pulse |
| Detector | InSb, room temperature–1600 °C |
| Samples | Round Ø6/8/10/12.7/20/25.4 mm; square 10 × 10/20 × 20 mm; thickness 0.1–6 mm |
| Positions | Three- or five-sample automatic changer |
| Atmosphere/vacuum | Inert gas or vacuum <2 × 10⁻⁵ mbar |