LFA 717 HyperFlash
Flash-method thermal diffusivity and heat capacity in seconds without damaging the sample.
LFA 717 HyperFlash measures 0.01–2000 mm²/s and reaches 1250 °C in the HT version.
LFA 717 HyperFlash delivers a short xenon-light pulse to one side of a specimen and records the temperature rise on the other. The response yields thermal diffusivity and, with heat capacity and density, conductivity. Standard and HT versions cover −100 to 500 °C and room temperature to 1250 °C respectively, with up to 16 or four sample positions.
| Instrument type | Light-flash thermal diffusivity/conductivity analyzer |
|---|---|
| Versions | HyperFlash −100–500 °C (room-temperature option); HT room temperature–1250 °C |
| Maximum heating rate | 50 K/min |
| Thermal diffusivity range | 0.01–2000 mm²/s |
| Thermal conductivity range | 0.1–4000 W/(m·K) |
| Diffusivity accuracy | ±3%; repeatability ±2% |
| IR detectors | InSb room temperature–500 °C and HT to 1250 °C; MCT −100–500 °C |
| Pulse source | Xenon lamp to 10 J/pulse, 10–1500 µs |
| Sample sizes | Round Ø6, 8, 10, 12.7 mm, maximum 25.4 mm; square; HT Ø10/12.7 mm or 10 mm square |
| Sample thickness | Typically 0.5–3 mm |
| Positions | Up to 16; four HT in Al₂O₃ mini-tube furnaces with individual thermocouples |
| Atmospheres | Inert/oxidizing, static/dynamic |
| Vacuum | <150 mbar; 10⁻⁴ mbar with turbopump on HT |
| Standards | ASTM E1461, E2585; ISO 22007-4, 18755; DIN EN 821-2; JIS R1611; GB/T22588 |