LOI 901 (oxygen index)
Find the minimum oxygen concentration that sustains combustion for flame-retardant development.
LOI 901 measures oxygen index from 0 to 100% with automatic calibration under ISO 4589-2.
LOI 901 is a benchtop limiting-oxygen-index analyzer for plastics. Two borosilicate columns accept rod, flat and round samples. A paramagnetic oxygen cell and automatic gas mixing control the O₂/N₂ atmosphere, while the touchscreen records ignition time and burning distance.
| Equipment type | Benchtop limiting oxygen index analyzer |
|---|---|
| Method | Room-temperature plastic flammability by oxygen index |
| Test columns | Two borosilicate glass columns, Ø75 × 450 mm |
| Rod sample | 6.5 × 3.0 × 70–150 mm |
| Flat sample | 52 × 140 mm, as-made thickness |
| Holders | Three for rod, flat and round specimens |
| Oxygen range | 0–100%; ±0.1% error |
| Ignition | Handheld propane igniter with fine and shutoff valves |
| Control | Single-board computer, 24-bit ADC, paramagnetic O₂ cell, solenoid/control valves for both gases; automatic calibration/test |
| Display | 7-inch 1024 × 800 touchscreen showing O₂, mixture, flow and time |
| Outputs | Oxygen index, minimum sustaining O₂, burn time and burn length |
| Dimensions (W×D×H) | 360 × 360 × 630 mm |
| Weight | 21 kg |
| Power | 110–230 V, 50/60 Hz, up to 40 W |
| Gases | O₂ ≥98% at 4.0 or 7.0 bar; N₂ ≥98% at 4.0 bar |
| Operating conditions | 23 ±5 °C, 50 ±20% relative humidity |
| Software | Windows 10 IoT, included LOI2016_SBC license |
| Standards | ISO 4589-2, ASTM D2863 |