NanoTR and PicoTR use time-domain thermoreflectance to analyze thin-film heat transport. NanoTR uses nanosecond excitation for thicker layers, while PicoTR resolves films from 10 nm in suitable materials. Front- or rear-heating and optional scanned maps support coatings, electronics and layered specimens.
| Instrument type | Time-domain thermoreflectance (TDTR) thin-film thermal analyzer |
|---|---|
| Models | NanoTR nanosecond; PicoTR picosecond |
| Heating laser | NanoTR 1 ns/1550 nm/100 µm spot; PicoTR 0.5 ps/1550 nm/45 µm |
| Probe laser | NanoTR continuous 785 nm/50 µm; PicoTR 0.5 ps/775 nm/25 µm |
| Diffusivity range | 0.01–1000 mm²/s |
| Accuracy | ±6.2% product page CRM; ±7.9% brochure for another reference |
| Repeatability | ±5% |
| Rear-heating film thickness | NanoTR: resin 30 nm–2 µm, ceramic 300 nm–5 µm, metal 1–20 µm; PicoTR: resin 10–100 nm, ceramic 10–300 nm, metal 100–900 nm |
| Front-heating film thickness | >1 µm NanoTR; >100 nm PicoTR |
| Substrate | 10–20 mm square, up to 1 mm thick, transparent or opaque |
| Temperature | Room temperature or to 500 °C; PicoTR also −100–500 °C |
| Measurement time | <30 s NanoTR; <5 min PicoTR |
| Mapping option | NanoTR 10 × 10 mm, 1 µm resolution; PicoTR 8 × 14 mm, 2 µm |
| Weight | 40 kg NanoTR; 90 kg PicoTR |
| Power | 100–240 V ±10%, 50/60 Hz; 0.5 kVA NanoTR, 1.5 kVA PicoTR |