Rosand RH2000
Capillary rheometry to 10⁸ s⁻¹ for extrusion, molding and spraying.
Rosand RH2000 is a benchtop high-pressure platform with single- or twin-bore versions.
Rosand RH2000 measures high-shear melt behavior beyond rotational rheometers. RH2100 and RH2200 provide one or two bores, with standard 12 kN force and optional 20 kN. Flowmaster applies Bagley, Rabinowitsch and Hagenbach corrections and the Cogswell extensional-viscosity model.
| Instrument type | Benchtop high-pressure capillary rheometer |
|---|---|
| Versions | RH2100 single bore; RH2200 twin bore |
| Maximum force | 12 kN standard, 20 kN optional; 100 kN frame stiffness |
| Maximum piston speed | 600 mm/min standard; 1200 high-speed |
| Speed dynamic range | 200,000:1; 400,000:1 high-speed; <0.1% error |
| Shear rate | To 10⁸ s⁻¹, die/bore dependent |
| Temperature | Room temperature–400 °C, optional 500; 5–300 °C with low-temperature coil |
| Temperature stability | Better than ±0.1 °C |
| Heating zones | Three independent |
| Bore diameter | 15 mm standard; optional 9.5/12/19/24 mm; 250 mm channel |
| Bore material | Nitrided steel; optional Hastelloy/stainless |
| Pressure transducers | 250–30,000 psi ranges; ±0.25% accuracy |
| Dies | Tungsten carbide ±5 µm; Ø0.5–2 mm in 0.5 mm steps and 3 mm; L/D 10–60; zero-length for extensional viscosity |
| Options | 1 µm-resolution laser extrudate swell, melt cutter, nitrogen purge, alternative dies/transducers |
| Dimensions | Rheometer 1.1 × 0.55 × 0.65 m H×W×D; electronics 0.65 × 0.34 × 0.53 m |
| Weight | 120 kg rheometer; 30 kg electronics |
| Power | Single-phase 230 V ±10%, 50–60 Hz, 16 A |
| Software | Flowmaster with Bagley/Rabinowitsch/Hagenbach corrections and Cogswell model |
| Standards | Rosand range: ISO 11443, 17744; ASTM D3835, D5099 |