Rosand RH7
50 kN twin-bore capillary rheometry for highly viscous and filled materials.
Rosand RH7 reaches 5 × 10⁷ s⁻¹ and offers die swell, melt strength and pVT options.
Rosand RH7 increases capillary rheometer force to 50 kN for compounds that exceed benchtop limits. Its twin bores and programmable piston measure melt flow at high shear rates. Options include laser extrudate swell, melt strength, slot dies and pVT testing.
| Instrument type | Floor-standing high-force twin-bore capillary rheometer |
|---|---|
| Bores | Two |
| Maximum force | 50 kN; 250 kN frame stiffness |
| Maximum piston speed | 600 mm/min |
| Speed dynamic range | To 400,000:1; <0.1% error |
| Shear rate | To 5 × 10⁷ s⁻¹ |
| Temperature | Room temperature–400 °C, optional 500; 5–300 °C with low-temp coil |
| Temperature stability | Better than ±0.1 °C; three heating zones |
| Bore diameter | 15 mm standard; optional 9.5/12/19/24 mm; 290 mm channel |
| Bore material | Nitrided steel; optional Hastelloy/stainless |
| Pressure transducers | 250–30,000 psi; ±0.25% accuracy |
| Dies | Tungsten carbide ±5 µm; Ø0.5–2 mm in 0.5 mm steps and 3 mm; zero-length dies |
| Options | Tragethon melt-strength take-up with 76/125 mm PTFE-coated drums; automatic winding; uni-/biaxial 1 µm laser swell; 0.5–2.5 mm slot dies with direct differential pressure; pVT, nitrogen purge, melt cutters |
| Dimensions | Height 2.45 m; width 715 mm transport/780 connected; depth 765 mm or 900 with take-up |
| Weight | 350 kg without accessories |
| Power | 3 × 230 V (3P+N+PE) ±10%, 3 × 16 A; UL/CSA 3 × 220 V (3P+PE), 3 × 20 A |
| Software | Flowmaster |
| Standards | Rosand range: ISO 11443, 17744; ASTM D3835, D5099 |