Test refractory behavior under operating temperature, sustained load and time.
RUL/CIC 421 measures refractoriness under load and compression creep to 1700 °C.
RUL/CIC 421 measures deformation of refractory specimens during heating under constant or changing compression. E/6 reaches 1700 °C, while G/6 offers vacuum and controlled gas atmospheres to 1600 °C. Its differential displacement system resolves 5 nm digitally for load-bearing and creep measurements.
| Instrument type | Refractoriness-under-load (RUL) and compression-creep (CIC) tester |
|---|---|
| Official name | RUL/CIC 421 |
| Versions | E/6 and G/6 |
| Temperature | Room temperature–1700 °C E/6; to 1600 °C G/6 |
| Load | 1–1000 N in 1–100 N steps; maximum stress 0.5 N/mm² |
| Variable-load option | Preload to 300 N; test load 0–700 N at 0.3–3 N/s |
| Specimen | Ø50 × 50 mm with Ø12.5 mm coaxial hole; alternative Ø36 mm to GOST 4070-2000 |
| Positions | One |
| Deformation range | 20 mm; 4,000,000 steps; 5 nm digital resolution |
| Measurement system | Differential, inner/outer corundum tubes |
| Heaters | Four Super-Kanthal 1800 elements |
| Heating rate | 5 K/min in manufacturer examples |
| Atmospheres | E/6 static air, optional inert purge; G/6 static/dynamic air/inert gas, vacuum to 10⁻² mbar |
| Software | Windows Proteus, temperature program to 96 steps |
| Standards | ISO 1893, 3187, 16835; DIN EN 993-8/-9; ASTM C832-00 |